FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 5171699
·
Received October 23, 2015
Report
- Report Number
- 2032227-2015-58744
- Event Type
- Malfunction
- Date Received
- October 23, 2015
- Date of Event
- October 3, 2015
- Report Date
- October 3, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
UNIT RECEIVED WITH BUTTON ERROR ALARM AND HAD INTERMITTENT ACT AND DOWN BUTTONS RESPONSE DUE TO CORRODED KEYPAD TRACES. UNIT HAD MINOR SCRATCHED LCD WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNER AND CRACKED RESERVOIR TUBE LIP. (B)(4).
Description of Event or Problem · 1
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED BUTTON ERROR AND THE KEYPAD BUTTONS ARE NOT RESPONSIVE. CUSTOMER DOES NOT RECALL ANY SIGNIFICANT EVENTS LEADING TO THE ERROR. CUSTOMER'S BLOOD GLUCOSE WAS 135 MG/DL UNKNOWN AT THE TIME OF INCIDENT. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO A BACK-UP PLAN PER DOCTOR'S INSTRUCTION. THE CUSTOMER WAS ADVISED THAT THE DEVICE WOULD BE REPLACED AND TO RETURN THE PRODUCT FOR ANALYSIS. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 700960 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR |