RADICAL-7
Report
- Report Number
- 2031172-2015-01189
- Event Type
- Malfunction
- Date Received
- October 21, 2015
- Date of Event
- September 16, 2015
- Report Date
- September 25, 2015
- Manufacturer
- MASIMO
- Product Code
- DQA
- PMA / PMN Number
- K110028
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
THE RETUNED DEVICE WAS EVALUATED. DURING INVESTIGATION IT WAS FOUND THAT POWERING ON FOLLOWING ABRUPT POWER OFF THE "ERROR IN DATA STORAGE" MESSAGE IS PRESENT. POWERING ON FOLLOWING STANDARD SHUTDOWN SEQUENCE THERE ARE NO ERRORS PRESENT. LOG FILE INDICATES 12 TIMES THE RADICAL-7 WAS ABRUPTLY POWERED OFF. BY DESIGN THE RADICAL-7 MAY DISPLAY THE "ERROR IN DATA STORAGE" MESSAGE WHEN POWERING ON AFTER AN ABRUPT SHUTDOWN. THIS ERROR OCCURS WHEN THE RADICAL-7 DOES NOT GET TO RUN THROUGH ITS STANDARD SHUTDOWN PROTOCOL; THE ERROR OCCURS WHEN THE POWER IS INTERRUPTED WHILE THE DEVICE IS WRITING INFORMATION TO THE DEVICE MEMORY. CUSTOMER COMPLAINT REGARDING SPO2 FLUCTUATION NOT CONFIRMED. COMPLAINT REGARDING "ERROR IN DATA STORAGE" CONFIRMED.
IT WAS REPORTED THAT THE SPO2 VALUE SHOWED DRASTICALLY FLUCTUATION 85-95% RANGE. SENSOR PLACEMENT WAS AT THE TOE. THE SITUATION WAS NOT CHANGED AFTER REPLACE THE CABLE AND SENSOR. THE SITUATION WAS NOT CHANGED AFTER SENSOR PLACEMENT AREA (TOE) WAS SWITCHED. THEN, THE SPO2 MEASUREMENT WAS SWITCHED TO NIHON KOHDEN MONITOR AND SPO2 VALUE 100% CAME UP. BLOOD GAS ANALYZER ALSO SHOWED 100%. PI: 1-3 THERE WERE NO LIGHT EMISSION OR/AND ELECTRICAL INSTRUMENTS AROUND SENSOR PLACEMENT AREA. TECH SUPPORT-MJC CONFIRMED "ERROR IN DATA STORAGE" MESSAGE CAME UP, OCCURS IN BOTH CONDITION OF THE AC DRIVE AND BATTERY DRIVE. THERE WAS NO CONSEQUENCE OR IMPACT TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 695104 | RADICAL-7 | OXIMETER | DQA | MASIMO | 23789 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |