FDA Adverse Event
Other
Summary report: N
ARCHITECT I2000 ANALYZER
MDR report key: 516388
·
Received January 23, 2004
Report
- Report Number
- 1628664-2004-00003
- Event Type
- Other
- Date Received
- January 23, 2004
- Date of Event
- December 22, 2003
- Report Date
- January 20, 2004
- Manufacturer
- ABBOTT MFG, INC.
- Product Code
- JJE
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE CUSTOMER STATES THAT AN INITIAL ARCHITECT BHCG RESULT OF 238.42 IU/L WAS GENERATED ON A PT. THE ACCOUNT HAS A POLICY OF RETESTING ANY BHCG RESULTS THAT FALL WITHIN THE RANGE OF 5 - 1000 IU/L. THE CUSTOMER THEREFORE RETESTED THE SAMPLE AND RESULTS OF 235.53 AND <1.20 IU/L WERE GENERATED. THE CUSTOMER STATES THAT OTHER REPRODUCTIVE HORMONES CONFIRMED LIKELY DIAGNOSIS OF PREGNANCY. THE CUSTOMER REPORTED THE RESULT OF 238.42 IU/L. NO IMPACT TO PT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ARCHITECT I2000 ANALYZER | AUTOMATED IMMUNOASSAY ANALYZER | JJE | ABBOTT MFG, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | ARCHITECT BHCG REAGENT LOT NO. 01577M100. |