FDA Adverse Event Injury Summary report: N

TALENT TAA STENT GRAFT

MDR report key: 5162836 · Received October 20, 2015

Report

Report Number
2953200-2015-01769
Event Type
Injury
Date Received
October 20, 2015
Date of Event
June 26, 2015
Report Date
September 25, 2015
Manufacturer
MEDTRONIC IRELAND
Product Code
MIH
PMA / PMN Number
P070007
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION CONCLUSION: OFF-LABEL, UNAPPROVED, OR CONTRAINDICATED USE (STENT GRAFT IMPLANTED IN VENA CAVA).

Description of Event or Problem · 1

MEDTRONIC RECEIVED THE FOLLOWING INFORMATION OBTAINED FROM THE JOURNAL ARTICLE ENTITLED; INFERIOR VENA CAVA STENT-GRAFT SEPSIS ERIC I. JENG, DAMARIS ORTIZ, ARSHAD KHAN, ENRICO BENEDETTI, MARTIN BORHANI, AND JOSE OBERHOLZER, (ANNALS OF VASCULAR SURGERY 2015;29 :1451.E17-1451.E20) A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN THE PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN IATROGENIC INFERIOR VENA CAVA (IVC) INJURY AS A LIFE-SAVING MEASURE. WE PRESENT THE CASE OF A PATIENT WITH A COMPLICATED MEDICAL HISTORY, WHICH INCLUDED STENT GRAFTING AS A LIFE-SAVING MEASURE FOR AN IATROGENIC INFERIOR VENA CAVA (IVC) INJURY. FOR PERSISTENT SEPSIS SECONDARY TO STENT-GRAFT INFECTION, THE PATIENT UNDERWENT EXTRACTION OF 2 IVC STENT GRAFTS, PRIMARY REPAIR OF A DUODENAL-CAVAL FISTULA, AND REPAIR OF THE IVC WITH AN ALLOGRAFT VEIN PATCH. DISCUSSION OF THIS CASE SOUGHT TO SHED LIGHT ON THE INTRICACIES INVOLVED IN MEDICAL DECISION MAKING IN AN ERA OF ADVANCED MEDICAL TECHNOLOGY. OUR PATIENT IS A (B)(6) WOMAN WITH A SURGICAL HISTORY SIGNIFICANT FOR AN ATTEMPTED RETROPERITONEOSCOPIC RIGHT ADRENALECTOMY PERFORMED AT AN OUTSIDE INSTITUTION FOR THE TREATMENT OF A SYMPTOMATIC NEOPLASM 3 YEARS EARLIER. HER INITIAL OPERATIVE COURSE WAS COMPLICATED BY SIGNIFICANT BLEEDING FROM THE RIGHT ADRENAL VEIN, THE RIGHT RENAL VEIN AND THE IVC. DESPITE CONVERTING TO LAPAROTOMY, AND PERFORMING A RIGHT ADRENALECTOMY AND RIGHT NEPHRECTOMY, THE OPERATING SURGEONS WERE UNABLE TO STOP THE BLEEDING. VASCULAR SURGERY WAS EMERGENTLY CONSULTED, AND AFTER PLEGETTED REPAIR OF SEVERAL AVULSED LUMBAR VEINS, IT WAS CLEAR THAT MOST BLEEDING STEMMED FROM THE RETROHEPATIC IVC. ALTERNATIVE APPROACHES TO CONTROL HEMORRHAGE AT THIS TIME INCLUDE AN ATRIOCAVAL SHUNT AS INITIALLY DESCRIBED BY (B)(6) IN 1968, COMPLETE IVC LIGATION, AND/OR ENDOVASCULAR MEANS, AS WAS CHOSEN. CITING POOR VISUALIZATION AND TECHNICAL DIFFICULTY, AN ENDOVASCULAR APPROACH WAS USED, STARTING WITH A RIGHT FEMORAL VEIN CUTDOWN. THROUGH A 12F SHEATH, OVER A BENSON WIRE, A RELIANT BALLOON WAS ADVANCED TO TEMPORARILY CONTROL HEMORRHAGE. ALTHOUGH THIS WAS UNSUCCESSFUL IN CONTROLLING THE BLEEDING, THE PRESENCE OF THE INFLATED RELIANT BALLOON CREATED A PATHWAY THAT ENABLED THE BENSON WIRE TO CROSS THE LESION. THE RELIANT BALLOON WAS THEN REMOVED, AND A MARKER PIGTAIL CATHETER WAS ADVANCED CEPHALAD TO THE INJURED SEGMENT. USING THE MARKED PIGTAIL TO SIZE THE IVC, THROUGH A 22F SHEATH, A 32 X 115-MM TALENT-COVERED STENT GRAFT WAS DEPLOYED TO COVER THE INJURY. POSTPLACEMENT VENOGRAM DEMONSTRATED PERSISTENT EXTRAVASATION SUPERIOR TO THE STENT GRAFT. FOR THAT REASON, A 34 X 158 MM TALENT STENT GRAFT WAS PLACED CEPHALAD, OVERLAPPING THE FIRST. REPEAT POSTPLACEMENT VENOGRAPHY DEMONSTRATED COMPLETE COVERAGE OF THE INJURED SEGMENT AND A PATENT IVC. DURING THIS OPERATION, THE PATIENT RECEIVED 90-L CRYSTALLOID, 6-L ALBUMIN, 146 UNITS PACKED RED BLOOD CELLS, 48 UNITS OF FRESH FROZEN PLASMA, 13 PACKS OF PLATELETS, AND 12 UNITS CRYOPRECIPITATE. HER POSTOPERATIVE COURSE WAS FURTHER COMPLICATED BY RESPIRATORY FAILURE REQUIRING PROLONGED VENTILATOR SUPPORT, ACUTE RENAL FAILURE DEMANDING RENAL REPLACEMENT THERAPY, AND ACUTE CHOLECYSTITIS WITH CHOLEDOCHOLITHIASIS NECESSITATING CHOLECYSTOSTOMY TUBE, AND ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY WITH SPHINCTEROTOMY. THE TREATING PHYSICIANS AT THE OUTSIDE INSTITUTION PROCEEDED WITH AN INTERVAL CHOLECYSTECTOMY AND A CHOLEDOCHODUODENOSTOMY. DESPITE LONG-TERM ANTIBIOTICS, THE NEXT YEAR AND A HALF WERE MARKED BY PERSISTENT EPISODES OF POLYMICROBIAL INFECTIONS, SECONDARY TO CHOLANGITIS AND INFECTED IVC STENT GRAFTS. GIVEN THE PATIENT'S DETERIORATING COURSE, SHE WAS TRANSFERRED TO OUR INSTITUTION FOR FURTHER CARE. IT WAS CLEAR THAT SHE NEEDED SURGICAL TREATMENT FOR BOTH RECURRENT CHOLANGITIS AND THE INFECTED IVC STENT GRAFTS. PREOPERATIVE WORK-UP INCLUDED A REPEAT CT SCAN FOR OPERATIVE PLANNING THAT HAD PATHOGNOMONIC FINDINGS OF AN INFECTED IVC STENT GRAFT. WE CHOSE TO STAGE OUR OPERATION, AND ELECTED TO PROCEED FIRST WITH A HEPATICOJEJUNOSTOMY WITH THE INTENT OF TREATING THE RECURRENT CHOLANGITIS, AND ALLOW TIME FOR NUTRITIONAL OPTIMIZATION. AFTER AN 8-WEEK INTERVAL, SHE UNDERWENT EXPLANTATION OF THE STENT GRAFTS. PREOPERATIVELY, WE PLACED A LEFT FEMORAL VEIN CENTRAL LINE AND A LARGE RIGHT INTERNAL JUGULAR CENTRAL LINE FOR POSSIBILITY OF VENOVENOUS BYPASS. AFTER AN EXTENSIVE LYSIS OF ADHESIONS, THE RIGHT LOBE OF THE LIVER WAS MOBILIZED TO EXPOSE THE IVC. WE IDENTIFIED THE RIGHT HEPATIC, MIDDLE HEPATIC, AND LEFT HEPATIC VEINS AND OBTAINED CONTROL OF THE HEPATIC HILUM FOR POSSIBLE VASCULAR EXCLUSION OF THE LIVER SHOULD THAT HAVE BECOME NECESSARY. DISTAL CONTROL OF THE IVC WAS PERFORMED AS WELL AS EXPOSURE OF THE INFRARENAL AORTA. THE DISSECTION WAS CARRIED CEPHALAD INTO THE IVC UNTIL IT WAS CLEAR THAT THE SWEEP OF THE DUODENUM WAS ATTACHED AND ADHERENT TO THE IVC. THIS WAS TAKEN DOWN, REVEALING A LARGE DEFECT IN THE IVC COMMUNICATING WITH A LARGE DEFECT IN THE DUODENUM. THIS WAS CLEARLY A DUODENAL CAVAL FISTULA, WITH THE DEFECT IN THE IVC MEASURING 5 CM IN DIAMETER, WITH THE STENT GRAFT VISIBLE WITHIN IT. PROXIMAL AND DISTAL CONTROL OF THE IVC WAS ACHIEVED, AND THE FIRST THORACIC STENT GRAFT WAS THEN EXTRACTED. THE LUMBAR VEINS HAD BEEN OVERSEWN PREVIOUSLY. THE LUMEN OF THE IVC WAS REMARKABLE FOR A LARGE AMOUNT OF CHRONIC APPEARING THROMBUS, AND THIS WAS EXTRACTED UNDER DIRECT VISION. THE INDWELLING LEFT FEMORAL CATHETER WAS EXCHANGED FOR A 7F SHEATH, AND THE BERENSTEIN CATHETER AND A GUIDING BENSON WIRE WERE PASSED INTO THE IVC, AND A VENOGRAM WAS PERFORMED. THIS WAS SIGNIFICANT FOR THE SECOND THORACIC STENT GRAFT (FIG. 2) WITHIN THE IVC, AND SHOWED THERE WAS NO FLOW THROUGH THE STENT GRAFT INTO THE ATRIUM. CEPHALAD TO THE AREA WHERE THE FIRST STENT GRAFT WAS VISUALIZED, CAREFUL FINGER FRACTURE WAS USED TO LYSE THE ADHESIONS WITHIN THE LUMEN OF THE STENT GRAFT TO AVOID FURTHER IVC INJURY FROM THE EXPOSED BARE WIRES, AND THE ENTIRE STENT GRAFT WAS EXTRACTED. THE CEPHALAD ASPECT OF THE STENT GRAFT HAD FORMED A WEB OF CHRONIC APPEARING THROMBUS OR SCAR TISSUE THAT WAS OBSTRUCTING OR NEARLY OBSTRUCTING THE LUMEN. COPIOUS FORWARD AND BACKBLEEDING AND A VEIN PATCH ANGIOPLASTY WAS COMPLETED. GIVEN THE SIGNIFICANT BOWEL WALL EDEMA, THE ABDOMEN WAS KEPT OPEN. SHE WAS EXTUBATED ON POSTOPERATIVE DAY 2, AND SKIN CLOSURE OF THE ABDOMEN WAS COMPLETED BY POSTOPERATIVE DAY 12. PATHOLOGY OF THE SPECIMEN SHOWED AN ATHEROMATOUS PLAQUE WITH LUMINAL ACUTE FIBRIN INFLAMMATORY THROMBUS THAT WAS CONSISTENT WITH INFECTED GRAFT CONTENTS WITH POSITIVE CULTURES FOR ESCHERICHIA COLI, CANDIDA ALBICANS, COAGULASE NEGATIVE STAPHYLOCOCCUS, AND ENTEROCOCCUS FAECIUM. THE PATIENT WAS PLACED ON A 6-WEEK COURSE OF INTRAVENOUS ANTIBIOTICS AND DISCHARGED TO ACUTE REHABILITATION ON POSTOPERATIVE DAY 16. IT HAS BEEN 2 YEARS NOW, AND THE PATIENT IS HOME AND DOING WELL. SHE IS NOW FOLLOWED WITH ANNUAL ULTRASOUND SURVEILLANCE OF THE IVC. MEDICAL HISTORY OF IATROGENIC INFERIOR VENA CAVA INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
692029 TALENT TAA STENT GRAFT SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH MEDTRONIC IRELAND

Patients

Seq Age Sex Outcome Treatment
1 00049 YR Required Intervention