FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 5162638 · Received October 20, 2015

Report

Report Number
1416980-2015-39426
Event Type
Malfunction
Date Received
October 20, 2015
Date of Event
February 19, 2015
Report Date
October 20, 2015
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED AND THE EVALUATION IS COMPLETE. DURING THE EVENT HISTORY LOG REVIEW, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED. UPON CONCLUSION OF THE INVESTIGATION, THE CAUSE OF THIS ISSUE WAS DETERMINED TO BE ONE OR MORE CYCLES ADVANCING TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2015 AT 10:24:08. DURING NIGHT DRAIN CYCLE THREE, THE PATIENT'S ULTRAFILTRATION READING WAS 1949ML, INDICATING THE HOME PATIENT DRAINED 1949ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1900ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
692218 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1