NOVA MAX LINK GLUCOSE MONITOR
Report
- Report Number
- 3004193489-2015-00140
- Event Type
- Malfunction
- Date Received
- October 20, 2015
- Date of Event
- September 23, 2015
- Report Date
- September 24, 2015
- Manufacturer
- NOVA BIOMEDICAL CORP.,
- Product Code
- NBW
- PMA / PMN Number
- K040603
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- PATIENT
Narratives
TEST STRIP LOT # 1020215050, EXPIRATION DATE: 2017/02/28, CONTROL SOLUTION LOT # 1030314175, CSR 82-127 MG/DL. THE CUSTOMER'S COMPLAINT IS CONFIRMED. FAILURE ANALYSIS OF THE RETURNED DEVICE REVEALED THAT THE NOVA MAX LINK GLUCOSE MONITORING DEVICE PERFORMED WITHIN SPECIFICATION. VISUAL AS WELL AS CHEMICAL EVALUATION (BY TESTING WITH CONTROL SOLUTION) OF THE RETURNED GLUCOSE TEST STRIPS REVEALED THE STRIPS TO BE COMPROMISED. THE ROOT CAUSE COULD NOT BE CONCLUSIVELY IDENTIFIED. A POTENTIAL CONTRIBUTORY ROOT CAUSE MAY BE RELATED TO EXPOSURE OF THE TEST STRIPS TO EXTREMES IN TEMPERATURE CONTRARY TO THE STORAGE AND HANDLING AS INDICATED IN LABEL COPY (IFU/PACKAGE INSERT) THIS IS FURTHER SUPPORTED BY THE RESULTS OF THE RETAIN STRIP TESTING WHICH INDICATES THE PERFORMANCE OF THE RETAIN STRIPS ARE WITHIN PRODUCT SPECIFICATION.
CONSUMER CALLED IN CONCERNED ABOUT THE DISCREPANCY IN HER BG READINGS BETWEEN HER NOVA MAX LINK AND HER ONE TOUCH. (B)(6) 2015 AT 5:20 PM BG 107MG/DL NML THE CONSUMER STATED, "...SHE FELT LOWER....", AT 5:21 PM BG 48 MG/DL ONE TOUCH. (B)(6) 2015 AT 11:16 AM BG 118 MG/DL NML THE CONSUMER STATED, "...SHE FELT LOWER....", AT 11:17 AM BG 56 ONE TOUCH. THE CONSUMER ALLEGES A CONTROL SOLUTION TEST WAS PERFORMED AS INSTRUCTED IN THE DIRECTIONS FOR USE ON (B)(6) 2015. HOWEVER, IT IS UNKNOWN IF THE RESULTS WERE IN OR OUT OF CONTROL RANGE. A CONTROL SOLUTION TEST WAS PERFORMED WHILE TROUBLESHOOTING WITH CUSTOMER SUPPORT SHOWING THE TEST STRIPS TO FALL OUT OF RANGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 693218 | NOVA MAX LINK GLUCOSE MONITOR | GLUCOSE MONITOR | NBW | NOVA BIOMEDICAL CORP., | N/A | 1020215050 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |