FDA Adverse Event Malfunction Summary report: N

DEXCOM G4 PLATINUM CONTINUOUS MONITORING SYSTEM DEXCOM SHARE SYSTEM

MDR report key: 5160498 · Received October 20, 2015

Report

Report Number
3004753838-2015-72408
Event Type
Malfunction
Date Received
October 20, 2015
Date of Event
August 4, 2015
Report Date
September 23, 2015
Manufacturer
DEXCOM, INC.
Product Code
MDS
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2015 TO CLAIM INTERMITTENT VIBRATION ON (B)(6) 2015. AT THE TIME OF CONTACT THE PATIENT DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION.

Description of Event or Problem · 1

THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. A REVIEW FO THE DOWNLOADED RECEIVER LOG DID NOT FIND ANY ERRORS RELATED TO THE CUSTOMER COMPLAINT. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION. THE COMPLAINT OF AN INTERMITTENT VIBRATION COULD NOT BE CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
692410 DEXCOM G4 PLATINUM CONTINUOUS MONITORING SYSTEM DEXCOM SHARE SYSTEM MDS MDS DEXCOM, INC. MT22495-PNK 5202019

Patients

Seq Age Sex Outcome Treatment
1 48 YR