FDA Adverse Event Malfunction Summary report: N

HANA

MDR report key: 5155768 · Received October 16, 2015

Report

Report Number
5155768
Event Type
Malfunction
Date Received
October 16, 2015
Date of Event
September 29, 2015
Report Date
September 29, 2015
Manufacturer
MIZUHO ORTHOPEDIC SYSTEMS, INC.
Product Code
JEB
Adverse Event
Yes
Report Source
User Facility report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

A MIZUHO OSI ORTHOPAEDIC SURGERY TABLE EXPERIENCED A CRACKED CONNECTION BOX. IT APPEARS THAT THE PLASTIC BOX THAT HOLDS THE CONNECTOR ON THE HANA TABLE EASILY BROKE. THE CRACKS SEEM LIKELY TO BE DUE TO THE LONG CABLE END CONNECTOR THAT ATTACHES TO BOX AND ACTS AS A LEVER APPLYING PRESSURE TO THE BOX. MANUFACTURER RESPONSE FOR ORTHOPAEDIC SURGERY TABLE, MIZUHO OSI HANA ORTHOPAEDIC SURGERY TABLE (PER SITE REPORTER): AS OF THIS REPORT, THE MANUFACTURER HAS NOT RESPONDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
685967 HANA ORTHOPAEDIC SURGERY TABLE JEB MIZUHO ORTHOPEDIC SYSTEMS, INC. HANA

Patients

Seq Age Sex Outcome Treatment
1 NO OTHER DEVICES USED