FDA Adverse Event
Malfunction
Summary report: N
HANA
MDR report key: 5155768
·
Received October 16, 2015
Report
- Report Number
- 5155768
- Event Type
- Malfunction
- Date Received
- October 16, 2015
- Date of Event
- September 29, 2015
- Report Date
- September 29, 2015
- Manufacturer
- MIZUHO ORTHOPEDIC SYSTEMS, INC.
- Product Code
- JEB
- Adverse Event
- Yes
- Report Source
- User Facility report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
A MIZUHO OSI ORTHOPAEDIC SURGERY TABLE EXPERIENCED A CRACKED CONNECTION BOX. IT APPEARS THAT THE PLASTIC BOX THAT HOLDS THE CONNECTOR ON THE HANA TABLE EASILY BROKE. THE CRACKS SEEM LIKELY TO BE DUE TO THE LONG CABLE END CONNECTOR THAT ATTACHES TO BOX AND ACTS AS A LEVER APPLYING PRESSURE TO THE BOX. MANUFACTURER RESPONSE FOR ORTHOPAEDIC SURGERY TABLE, MIZUHO OSI HANA ORTHOPAEDIC SURGERY TABLE (PER SITE REPORTER): AS OF THIS REPORT, THE MANUFACTURER HAS NOT RESPONDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 685967 | HANA | ORTHOPAEDIC SURGERY TABLE | JEB | MIZUHO ORTHOPEDIC SYSTEMS, INC. | HANA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NO OTHER DEVICES USED |