MPACT ACETABULAR PE LINER Ø 36
Report
- Report Number
- 3005180920-2015-00240
- Event Type
- Malfunction
- Date Received
- October 15, 2015
- Date of Event
- September 16, 2015
- Report Date
- November 19, 2015
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LPH
- PMA / PMN Number
- K103721
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
ON 29 OCTOBER 2015 THE R&D JOINTS DIRECTOR MADE THE FOLLOWING ANALYSIS ON THE RETRIEVED IMPLANTS: THE METALLIC SHELL AND THE PE LINER HAVE BEEN VISUALLY INSPECTED. THERE IS NOTHING TO BE REPORTED. WITH THE DEDICATED INSTRUMENT THE PE LINER CAN BE EASILY FULLY INSERTED. ON 19 NOVEMBER 2015 IT WAS PREPARED A FINAL REPORT WITH THE INFORMATION ALREADY SUBMITTED IN THE INITIAL REPORT AND HERE ABOVE. ON THE SAME DATE THE FINAL REPORT WAS SENT TO THE INITIAL REPORTER AND THE CASE WAS CLOSED.
BATCH REVIEW PERFORMED ON 14 OCTOBER 2015: LOT 140553: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 10 APRIL 2014. EXPIRATION DATE: 2019-02-28. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. MPACT ACETABULAR SHELL Ø56 TWO-HOLES REF. 01.32.156DH LOT. 135868 (K132879): LOT 135868: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 19 MY 2014. EXPIRATION DATE: 2019-04-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. NOT RECEIVED YET.
LINER WOULD NOT FULLY SEAT IN THE SHELL. AFTER MULTIPLE ATTEMPTS TO SEAT THE LINER IT REMAINED PROUD BY 1MM. THE SURGERY WAS DELAYED BY 10-15 MINUTES. THE SURGEON REPLACED BOTH THE CUP AND LINER THAT FIT CORRECTLY. THE EXPLANTS WILL BE RETURNED. THERE ARE NO X-RAYS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 682576 | MPACT ACETABULAR PE LINER Ø 36 | ACETABULAR PE LINER | LPH | MEDACTA INTERNATIONAL SA | 140553 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR | Other |