FDA Adverse Event
Malfunction
Summary report: N
OT VERIO IQ METER
MDR report key: 5152434
·
Received October 15, 2015
Report
- Report Number
- 2939301-2015-44023
- Event Type
- Malfunction
- Date Received
- October 15, 2015
- Report Date
- October 1, 2015
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
Description of Event or Problem · 1
ON (B)(6) 2015, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING AN INACCURATE HIGH RESULT OF "73 MG/DL" ON THE SUBJECT METER COMPARED TO "51 MG/DL" ON ANOTHER METER, PERFORMED WITHIN 30 MINUTES OF EACH OTHER. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED RESULTS DID NOT MEET LIFESCAN'S ACCURACY CRITERIA AND THE ALLEGED INACCURACY ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 685415 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3779296 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR |