FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 5151801 · Received October 15, 2015

Report

Report Number
2032227-2015-56939
Event Type
Malfunction
Date Received
October 15, 2015
Date of Event
September 29, 2015
Report Date
October 13, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE UNIT HAD A BUTTON ERROR ALARM AND AN INTERMITTENT BUTTON RESPONSE DUE TO A CORRODED KEYPAD. THE UNIT HAD A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, A MINOR SCRATCHED LCD WINDOW, A SCRATCHED RESERVOIR TUBE WINDOW, A CRACKED RESERVOIR TUBE, AND A CRACKED RESERVOIR TUBE WINDOW.

Description of Event or Problem · 1

THE CUSTOMER CALLED IN TO REPORT THAT THE INSULIN PUMP ALARMED BUTTON ERROR. THE CUSTOMER COULD NOT RECALL ANY SIGNIFICANT EVENTS LEADING TO THE ISSUE. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS UNKNOWN. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. THE CUSTOMER WAS INFORMED THAT THE DEVICE WOULD BE REPLACED, AND WAS INFORMED TO RETURN THE MALFUNCTIONING DEVICE BACK FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
682123 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAH

Patients

Seq Age Sex Outcome Treatment
1 24 YR