FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 5151801
·
Received October 15, 2015
Report
- Report Number
- 2032227-2015-56939
- Event Type
- Malfunction
- Date Received
- October 15, 2015
- Date of Event
- September 29, 2015
- Report Date
- October 13, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE UNIT HAD A BUTTON ERROR ALARM AND AN INTERMITTENT BUTTON RESPONSE DUE TO A CORRODED KEYPAD. THE UNIT HAD A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, A MINOR SCRATCHED LCD WINDOW, A SCRATCHED RESERVOIR TUBE WINDOW, A CRACKED RESERVOIR TUBE, AND A CRACKED RESERVOIR TUBE WINDOW.
Description of Event or Problem · 1
THE CUSTOMER CALLED IN TO REPORT THAT THE INSULIN PUMP ALARMED BUTTON ERROR. THE CUSTOMER COULD NOT RECALL ANY SIGNIFICANT EVENTS LEADING TO THE ISSUE. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS UNKNOWN. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. THE CUSTOMER WAS INFORMED THAT THE DEVICE WOULD BE REPLACED, AND WAS INFORMED TO RETURN THE MALFUNCTIONING DEVICE BACK FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 682123 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 24 YR |