FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 5151798 · Received October 15, 2015

Report

Report Number
2032227-2015-56996
Event Type
Malfunction
Date Received
October 15, 2015
Date of Event
August 14, 2015
Report Date
August 14, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

PUMP RECEIVED WITH BLANK DISPLAY DUE TO CORROSION ON ELECTRONICS. UNABLE TO VERIFY LOW BATTERY LIFE AND BUTTON KEYPAD ANOMALY DUE TO BLANK DISPLAY. PUMP RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW, CRACKED BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP. (B)(4). THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP WAS COMPLETELY DEAD AND THE KEYPAD BUTTONS ARE NOT RESPONSIVE. CUSTOMER CHANGED THE BATTERY BUT THE PROBLEM PERSISTS. CUSTOMER'S BLOOD GLUCOSE WAS UNKNOWN AT THE TIME OF INCIDENT. THE CUSTOMER WAS ADVISED THAT THE DEVICE WOULD BE REPLACED AND TO RETURN THE PRODUCT FOR ANALYSIS.