FDA Adverse Event Injury Summary report: N

EON MINI IPG, 16-CHANNEL RECHARGEABLE

MDR report key: 5150266 · Received October 14, 2015

Report

Report Number
1627487-2015-10281
Event Type
Injury
Date Received
October 14, 2015
Date of Event
September 18, 2015
Report Date
December 3, 2015
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
GZB
PMA / PMN Number
P010032
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Additional Manufacturer Narrative · 1

SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.(B)(4)

Description of Event or Problem · 1

FOLLOW UP INFORMATION IDENTIFIED THE PATIENT UNDERWENT SURGICAL INTERVENTION WHERE HER IPG WAS REPLACED WITH A NEW ONE, WHICH RESOLVED THE ISSUES.

Description of Event or Problem · 1

IT WAS REPORTED THE PATIENT STARTED EXPERIENCING PAIN AND WARMING AT THE IPG SITE WHILE CHARGING AFTER SHE FELL OFF A LADDER AND LANDED DIRECTLY ON HER IPG. SURGICAL INTERVENTION MAY BE PENDING TO ADDRESS THE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
678919 EON MINI IPG, 16-CHANNEL RECHARGEABLE SCS IPG GZB ST. JUDE MEDICAL - NEUROMODULATION 3788 3612449

Patients

Seq Age Sex Outcome Treatment
1 67 YR Other