FDA Adverse Event
Injury
Summary report: N
COVIDIEN
MDR report key: 5145184
·
Received September 9, 2015
Report
- Report Number
- 5145184
- Event Type
- Injury
- Date Received
- September 9, 2015
- Date of Event
- June 15, 2015
- Report Date
- September 8, 2015
- Manufacturer
- COVIDEN / MEDTRONIC
- Product Code
- GDW
- Adverse Event
- Yes
- Report Source
- User Facility report
- Reporter Location
- HI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Description of Event or Problem · 1
PATIENT HAD A LAPAROSCOPIC APPENDECTOMY ON (B)(6) 2015. HE REQUIRED ADDITIONAL PROCEDURE (AT ANOTHER FACILITY ON THE MAINLAND) DUE TO COMPLICATION FROM THE LAPAROSCOPIC APPENDECTOMY. IT WAS FOUND THAT ONE OF THE FIRED STAPLES, WAS FOUND LEFT OPEN (WITHOUT THE TIPS CURLED AROUND TISSUE). THE RAW TIP OF THE STAPLE HAD GOTTEN CAUGHT ON THE TERMINAL ILEUM AND CAUSED AN OBSTRUCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 597979 | COVIDIEN | ENDO GIA 45 TAN / ENDO GIA 60 PURPLE | GDW | COVIDEN / MEDTRONIC | 45-TAN EGI45AVM | 60-PURPLE EGIA60AMT |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 14 YR | Hospitalization| R |