FDA Adverse Event Malfunction Summary report: N

DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 5141313 · Received October 9, 2015

Report

Report Number
3004753838-2015-65975
Event Type
Malfunction
Date Received
October 9, 2015
Date of Event
September 12, 2015
Report Date
September 12, 2015
Manufacturer
DEXCOM, INC.
Product Code
MDS
UDI-DI
00386270000019
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). IT WAS REPORTED THAT CALIBRATION WAS NOT PERFORMED CORRECTLY. THE DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM USER'S GUIDE STATES: DO NOT CALIBRATE WHEN YOUR RECEIVER SCREEN IS SHOWING THE RISING SINGLE ARROW OR DOUBLE ARROW. ALSO, DO NOT CALIBRATE WHEN YOUR RECEIVER SCREEN IS SHOWING THE FALLING SINGLE ARROW OR DOUBLE ARROW. CALIBRATING DURING SIGNIFICANT RISE OR FALL OF BLOOD GLUCOSE MAY AFFECT ACCURACY OF SENSOR GLUCOSE READINGS. DO NOT CALIBRATE IF THE GLUCOSE READING ERROR SYMBOL SHOWS IN THE STATUS AREA. DO NOT CALIBRATE IF THE OUT OF RANGE SYMBOL SHOWS IN THE STATUS AREA. ADDITIONALLY, AN EXPIRED SENSOR WAS USED. THE DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM USER'S GUIDE STATES: DO NOT INSERT SENSORS PAST THE EXPIRATION DATE. THE EXPIRATION DATE FORMAT IS YYYY-MM-DD. INSERT SENSORS ON OR BEFORE THE END OF THE CALENDAR DAY PRINTED ON THE SENSOR PACKAGE LABEL.

Description of Event or Problem · 1

PATIENT'S HUSBAND CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2015 TO REPORT CONTINUOUS GLUCOSE MONITORING (CGM) INACCURACIES COMPARED TO BLOOD GLUCOSE (BG) METER THAT OCCURRED ON (B)(6) 2015. THE SENSOR WAS INSERTED ON (B)(6) 2015. PATIENT USED AN EXPIRED SENSOR. CALIBRATION WAS NOT PERFORMED CORRECTLY. AT THE TIME OF CONTACT, NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
671233 DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM MDS MDS DEXCOM, INC. 9500-27 5191511 00386270000019

Patients

Seq Age Sex Outcome Treatment
1 76 YR