PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2015-55568
- Event Type
- Malfunction
- Date Received
- October 9, 2015
- Date of Event
- September 24, 2015
- Report Date
- September 24, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA
- Reporter Occupation
- PATIENT
Narratives
THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. THE PUMP WAS RECEIVED WITH OPERATING CURRENTS WITHIN SPECS. THE PUMP PASSED SELF-TEST, ALARM ERROR TEST AND DISPLACEMENT TEST. HOWEVER, THE PUMP WAS UNABLE TO PRIME DURING THE PRIME TEST AND MOTOR ERROR ALARM DURING BASIC OCCLUSION TEST DUE TO FAULTY FORCE SENSOR SYSTEM. THE PUMP WAS RECEIVED WITH BLEEDING GLASS ON LCD BOARD, SCRATCHES ON LCD WINDOW, AND CRACKED BATTERY TUBE THREADS. (B)(4).
THE CUSTOMER REPORTED VIA PHONE CALL OF A BUTTON ERROR ALARM FROM THE INSULIN PUMP. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS UNKNOWN. THE CUSTOMER STATED THAT SHE WENT TO WASH HER FACE AND THE PUMP FELL OFF. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND TO REVERT TO A BACKUP PLAN PER HEALTH CARE PROVIDER'S INSTRUCTIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 671098 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR |