FDA Adverse Event Malfunction Summary report: N

TANDEM T:SLIM INSULIN DELIVERY SYSTEM

MDR report key: 5140724 · Received October 9, 2015

Report

Report Number
3007981285-2015-55288
Event Type
Malfunction
Date Received
October 9, 2015
Date of Event
September 14, 2015
Report Date
September 14, 2015
Product Code
LZG
PMA / PMN Number
K111210
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. H3 OTHER TEXT : DEVICE NOT RETURNED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A LOW BATTERY POWER ALERT BECAME FROZEN ON THE PUMP SCREEN AND THE CUSTOMER WAS UNABLE TO CLEAR THE ALERT. DURING TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT, THE ALERT WAS ABLE TO BE CLEARED. THERE WAS NO IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
671639 TANDEM T:SLIM INSULIN DELIVERY SYSTEM INSULIN PUMP LZG 004628

Patients

Seq Age Sex Outcome Treatment
1 13 YR