2.4MM TI VA LOCKING SCREW STARDRIVE 18MM-STERILE
Report
- Report Number
- 2520274-2015-16512
- Event Type
- Injury
- Date Received
- October 9, 2015
- Date of Event
- March 17, 2015
- Report Date
- September 28, 2015
- Manufacturer
- SYNTHES USA
- Product Code
- HRS
- PMA / PMN Number
- PK102694
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. DEVICE IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. SUBJECT DEVICE HAS NOT BEEN RECEIVED. WITHOUT A LOT NUMBER, THE DEVICE HISTORY RECORD REVIEW AND THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
(B)(4). SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. INVESTIGATION IS ONGOING; NO CONCLUSION COULD BE DRAWN AS PRODUCT IS ENTERING THE COMPLAINT SYSTEM. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
PRODUCT RETURN WAS REPORTED ON THE FIRST SUPPLEMENTAL MEDWATCH. DATE OF RECEIPT ADDED TO THE APPROPRIATE FIELD. (B)(6). (B)(4). PRODUCT INVESTIGATION SUMMARY: RECEIVED PARTS: 1 X 04.111.620S / VA-LCP-2-COLUMN DRP2.4 VOLAR R SHAFT 2HO LOT NUMBER: 8890096, 1 X 04.210.120S / VA LOCKSCR Ø2.4 SELF-TAP L20 TAN LOT NUMBER: UNKNOWN, 2 X 04.210.118S / VA LOCKSCR Ø2.4 SELF-TAP L18 TAN LOT NUMBER: UNKNOWN, 3 X UNKNOWN SCREWS. THE PLATE IN QUESTION SHOWS WEAR AND TEAR SIGNS AND SCRATCHES ALLOVER ON THE SURFACE. THREE (3) SCREWS WERE BROKEN BETWEEN THE SCREW HEAD AND THE THREADED SHAFT. THE TORX-RECESSES OF ALL THREE SCREWS WERE INTACT. THE REMAINDER OF THE RECEIVED SCREWS WAS INTACT, BUT SHOWED SIGNS OF WEAR AND TEAR. THE MANUFACTURING REVIEW OF THE ARTICLES (WITH THE KNOWN LOT NUMBERS) SHOWS THAT THE PRODUCTION PROCEDURES WERE ACCORDING TO THE SPECIFICATIONS AND THERE WERE NO ISSUES THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. WITHOUT THE ORIGINAL LABEL, IT IS IMPOSSIBLE TO DISTINGUISH THE RECEIVED SCREWS. FURTHERMORE, THERE IS NO INFORMATION ABOUT THE CORRESPONDING LOT NUMBERS OF THE SCREWS. THE MICROSCOPIC VIEWS OF THE BROKEN SURFACES SHOW A HOMOGENOUS SURFACE, WHICH INDICATES MATERIAL CONFORMITY. THE MANUFACTURER IS NOT AWARE OF ANY QUALITY PROBLEMS OR FAILURES CAUSED BY A FAULTY PRODUCT ON THE ARTICLE. ALSO, THERE ARE NOT ANY OTHER COMPLAINTS FOR THIS ARTICLE. BECAUSE OF THE DAMAGE TO THE SCREWS, THE COMPLAINT RELEVANT DIMENSIONS CANNOT BE CHECKED FOR DIMENSIONAL ACCURACY OF THE VALID MANUFACTURING SPECIFICATIONS. BASED ON THE PROVIDED CLINICAL INFORMATION AND THE INVESTIGATION RESULTS, AN EXACT ROOT CAUSE CANNOT BE DETERMINED. MULTIPLE FACTORS CAN LEAD TO TECHNICAL DIFFICULTIES DURING REMOVAL. THESE FACTORS INCLUDE, BUT ARE NOT LIMITED TO: WRONG TORQUE OR ANGULATION OF THE SCREWS DURING INSERTION, OR PROTEINS WHICH ARE FORMING A BOND BETWEEN THE PLATE AND SCREW THREAD. SOME OF THESE FACTORS ARE PATIENT SPECIFIC FACTORS OR THEY DEPEND ON THE PROPER CARE OF THE SURGEON, AND CAN THUS NOT BE INVESTIGATED NOR CONTROLLED BY DEPUY SYNTHES. IT IS CONCLUDED THAT THE CAUSE OF FAILURE IS NOT DUE TO ANY MANUFACTURING NON-CONFORMANCES. ORIGINAL IMPLANT PROCEDURE TOOK PLACE ON AN UNKNOWN DATE IN (B)(6) 2014. DEVICE USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT A RECONSTRUCTION SURGERY TOOK PLACE ON (B)(6) 2015 FOR A PATIENT BEING RUPTURED IN HER FLEXOR TENDON AROUND PLATE IMPLANTATION WHICH TREATED IN (B)(6) 2014. DURING SURGERY OF REMOVING IMPLANTS, THE SURGEON DETECTED THE BREAKAGE OF REPORTED SCREWS AT THEIR HEAD (2X 04.210.118S AND 1X 04.210.120S). ACCORDINGLY HE REMOVED THE SCREWS AND PLATE AND COMPLETED THE RECONSTRUCTION SURGERY SUCCESSFULLY. A PATIENT IS NOW UNDER OBSERVATION. THIS REPORT IS 3 OF 4 FOR (B)(4).
CLARIFICATION: PATIENT SUFFERED FLEXOR MUSCLE TENDON TEARING SOMETIME AFTER THE INITIAL IMPLANT PROCEDURE IN (B)(6) 2014.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 671777 | 2.4MM TI VA LOCKING SCREW STARDRIVE 18MM-STERILE | PLATE,FIXATION,BONE | HRS | SYNTHES USA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |