FDA Adverse Event
Other
Summary report: N
HANAUPORT
MDR report key: 5137561
·
Received July 27, 2004
Report
- Report Number
- 3004147784-2004-00002
- Event Type
- Other
- Date Received
- July 27, 2004
- Report Date
- December 10, 2003
- Manufacturer
- HERAEUS MEDICAL, INC.
- Product Code
- FRC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
IT WAS REPORTED THAT DURING THE CLEANING OF ROOM 8, A PIECE OF EQUIPMENT WAS UNPLUGGED FROM AN ELECTRICAL OUTLET IN THE COLUMN WHEN A SPARK WAS SEEN AND A FIRE ENSUED. GETINGE USA REPRESENTATIVES EXAMINED THE COLUMN NOTING THAT THE FIRE APPEARS TO BE CAUSED BY THE COMBINATION OF A LOOSE OXYGEN HOSE (CREATING OXYGEN RICH ENVIRONMENT) AND THE SPARK GENERATED WHEN A TECHNICIAN UNPLUGGED EQUIPMENT THAT STILL HAD CURRENT RUNNING TO IT.
Description of Event or Problem · 1
UNIT REPORTED TO HAVE A FIRE WITH THE ENCLOSURE. HEAT ASSOCIATED WITH THE FIRE CAUSED DAMAGE TO AN OUTLET, ELECTRICAL CORD AND GAS SUPPLY TUBING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HANAUPORT | GAS AND POWER DISTRIBUTION COLUMN | FRC | HERAEUS MEDICAL, INC. | 0007-5251 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |