Additional Manufacturer Narrative · 1

IT WAS REPORTED THAT DURING THE CLEANING OF ROOM 8, A PIECE OF EQUIPMENT WAS UNPLUGGED FROM AN ELECTRICAL OUTLET IN THE COLUMN WHEN A SPARK WAS SEEN AND A FIRE ENSUED. GETINGE USA REPRESENTATIVES EXAMINED THE COLUMN NOTING THAT THE FIRE APPEARS TO BE CAUSED BY THE COMBINATION OF A LOOSE OXYGEN HOSE (CREATING OXYGEN RICH ENVIRONMENT) AND THE SPARK GENERATED WHEN A TECHNICIAN UNPLUGGED EQUIPMENT THAT STILL HAD CURRENT RUNNING TO IT.

Description of Event or Problem · 1

UNIT REPORTED TO HAVE A FIRE WITH THE ENCLOSURE. HEAT ASSOCIATED WITH THE FIRE CAUSED DAMAGE TO AN OUTLET, ELECTRICAL CORD AND GAS SUPPLY TUBING.