ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2015-36021
- Event Type
- Malfunction
- Date Received
- October 7, 2015
- Report Date
- September 17, 2015
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP #1: DATE OF SUBMISSION 11/13/2015 DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2015 WITH THE FOLLOWING FINDINGS: THE BLACK BOX SHOWED EVIDENCE OF SEVERAL MANUAL TIME/DATE CHANGES. NO UNEXPLAINED TIME LOSS/GAIN WAS OBSERVED. THE ALLEGED ISSUE COULD NOT BE DUPLICATED.
THE PUMP HAS BEEN RETURNED TO ANIMAS BUT NOT YET EVALUATED. WHEN THE DEVICE EVALUATION IS COMPLETE, A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE PUMP'S CLOCK WAS INACCURATE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 663912 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 14 YR |