530G INSULIN PUMP

MDR report key: 5132553 · Received October 7, 2015

Report

Report Number
2032227-2015-53952
Event Type
Malfunction
Date Received
October 7, 2015
Date of Event
September 20, 2015
Report Date
September 20, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH BLANK DISPLAY DUE TO INTERFACE BOARD BROKEN SOLDER JOINT. THE REWIND, BASIC OCCLUSION, OCCLUSION, PRIMING, EXCESSIVE NO DELIVERY ALARM AND DISPLACEMENT TESTS WERE UNABLE TO BE PERFORMED DUE TO BLANK DISPLAY ANOMALY. THE DEVICE HAS MINOR SCRATCHES ON THE LCD WINDOW, CRACKED CASE NEAR DISPLAY WINDOW CORNERS AND CRACKED RESERVOIR TUBE LIP. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED VIA PHONE CALL THAT THEY HAVE A BLANK SCREEN ON THEIR INSULIN PUMP. CUSTOMER ADVISED THEY WOKE UP WITH LOW BLOOD GLUCOSE IN THE MORNING AND MANUALLY SUSPENDED THE DEVICE WHEN THEY REALIZED THE SCREEN WAS BLANK. TROUBLESHOOTING FOR THE BLANK DISPLAY WAS PERFORMED. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO A BACKUP PLAN. THE CUSTOMER WAS ADVISED THAT THE DEVICE WOULD BE REPLACED AND AGREED TO RETURN THE PRODUCT FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
663356 530G INSULIN PUMP OZO OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-551NAL

Patients

Seq Age Sex Outcome Treatment
1 48 YR