BLOOD GLUCOSE SYSTEM
Doing adverse event research by hand?
The Research Assistant searches MAUDE for you, groups reports by failure mode, and cites every record.
Free to try · every answer cites its records
Report
- Report Number
- 1052693-2015-01920
- Event Type
- Malfunction
- Date Received
- October 7, 2015
- Date of Event
- September 16, 2015
- Report Date
- June 27, 2016
- Manufacturer
- NIPRO DIAGNOSTICS, INC.
- Product Code
- NBW
- PMA / PMN Number
- K080641
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC
- Reporter Occupation
- OTHER
Narratives
(B)(4). PRODUCT NOT YET RETURNED FOR EVALUATION.
(B)(4). RETURNED METER AND RETURNED TEST STRIPS EVALUATED WITH NO DEFECTS FOUND. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: STRIP ISSUE.
CUSTOMER COMPLAINS OF HI RESULTS. CUSTOMER STATES SHE FEELS WELL AND DOES NOT REQUIRE ANY MEDICAL ATTENTION AT THIS TIME. THE CUSTOMER'S EXPECTED BLOOD RESULTS ARE 146-227MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 05/31/2018. CUSTOMER COULD NOT CONFIRM WHEN THE TEST STRIPS WERE FIRST OPENED BUT STATES THE STRIPS WERE PROPERLY STORED. CUSTOMER PERFORMED A BACK TO BACK BLOOD TEST HI AND HI NOT FASTING. REVIEWED METER MEMORY: (B)(6). CUSTOMERS CONCERN:ABOUT ALL THE RESULTS. NO ADVERSE EVENT REPORTED.
CUSTOMER COMPLAINS OF HI RESULTS. CUSTOMER STATES SHE FEELS WELL AND DOES NOT REQUIRE ANY MEDICAL ATTENTION AT THIS TIME. THE CUSTOMER'S EXPECTED BLOOD RESULTS ARE 146-227MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 05/31/2018. CUSTOMER COULD NOT CONFIRM WHEN THE TEST STRIPS WERE FIRST OPENED BUT STATES THE STRIPS WERE PROPERLY STORED. CUSTOMER PERFORMED A BACK TO BACK BLOOD TEST HI AND HI NOT FASTING. REVIEWED METER MEMORY: HI (B)(6) 2015 04:12:00 AM FASTING:YES; THE 368MG/DL (B)(6) 2015 01:46:00 AM FASTING:YES; THE 582MG/DL (B)(6) 2015 09:48:00 AM FASTING:YES; THE HI (B)(6) 2015 09:46:00 AM FASTING:YES; THE 257MG/DL (B)(6) 2015 03:03:00 AM FASTING:YES. CUSTOMERS CONCERN:ABOUT ALL THE RESULTS. NO ADVERSE EVENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 662526 | BLOOD GLUCOSE SYSTEM | NBW | NIPRO DIAGNOSTICS, INC. | TRUERESULT | PS2365 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |