FDA Adverse Event Malfunction Summary report: N

LAMINA SPREADER F/T-PAL

MDR report key: 5131148 · Received October 7, 2015

Report

Report Number
9680938-2015-10095
Event Type
Malfunction
Date Received
October 7, 2015
Report Date
September 25, 2015
Manufacturer
SYNTHES TUTTLINGEN
Product Code
GEN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED, THE INVESTIGATION COULD NOT BE COMPLETED, AND NO CONCLUSION COULD BE DRAWN, AS PRODUCT IS ENTERING THE COMPLAINT SYSTEM. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

A MANUFACTURING EVALUATION WAS COMPLETED: THE BOTTOM SPRING BETWEEN THE HANDLES HAS SHEARED OFF. THE MANUFACTURING EVALUATION SHOWS THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. THE SPRINGS ARE PRE BENT TO 15° BEFORE HARDENING AND THEN REQUIRE FURTHER BENDING AT FINAL ASSEMBLY TO REACH THE APPROPRIATE TENSION FOR THE JAWS TO CLOSE WHEN THE HANDLES ARE IN THE OPEN POSITION. A HARDNESS TEST WAS PERFORMED ON THE BROKEN SPRING AND FOUND TO BE CONFORMING. A MATERIAL ANALYSIS OF THE SPRING WAS CONDUCTED INTERNALLY AND CONFIRMED THAT THE MATERIAL IS CORRESPONDING TO THE DELIVERED MATERIAL CERTIFICATE. ALSO THE DIMENSIONS OF THE SPRING ARE CONFORMING. NO MANUFACTURING ISSUE WAS FOUND DURING INVESTIGATION. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. DEVICE IS NOT DISTRIBUTED IN THE UNITED STATES, BUT IS SIMILAR TO DEVICE MARKETED IN THE USA. DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. (B)(6). A DEVICE HISTORY RECORD REVIEW WAS PERFORMED FOR THE SUBJECT DEVICE LOT. THE REVIEW SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. THE RAW MATERIAL WHICH WAS DELIVERED AS LOT #864253 IS CORRESPONDING TO THE SPECIFICATIONS. THE HARDNESS WAS MEASURED AT THE TIME OF THE MANUFACTURING AT 48,5 HRC AND WAS FOUND TO BE GOOD. NO NCRS WERE GENERATED DURING PRODUCTION. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES EUROPE REPORTED AN EVENT (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THE SPRING ON THE HANDLE IS BROKEN. THE REPORTED ISSUE WAS DISCOVERED DURING SURGERY BUT BEFORE USE ON THE PATIENT. THERE WAS NO SURGICAL DELAY DUE TO THE REPORTED EVENT. THIS REPORT IS 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
665063 LAMINA SPREADER F/T-PAL FORCEPS, GEN & PLASTIC SURGERY GEN SYNTHES TUTTLINGEN T104531

Patients

Seq Age Sex Outcome Treatment
1