FDA Adverse Event Malfunction Summary report: N

ZIMMER MILL

MDR report key: 5130890 · Received October 2, 2015

Report

Report Number
2025102-2015-00004
Event Type
Malfunction
Date Received
October 2, 2015
Date of Event
August 31, 2015
Report Date
October 2, 2015
Manufacturer
BRASSELER USA, MEDICAL L.L.C.
Product Code
HSZ
PMA / PMN Number
K935367
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS HANDPIECE WAS SHIPPED 07/31/2014 AND HAS NOT BEEN RECEIVED BACK FOR SERVICING/PREVENTIVE MAINTENANCE. THE INFORMATION PROVIDED WITH THE COMPLAINT INDICATES THAT THE BUR WAS NOT PROPERLY LOCKED IN THE COLLET PRIOR TO OPERATING THE HANDPIECE. IF THE BUR IS NOT LOCKED IN PLACE THIS ALLOWS THE BUR TO MIGRATE DURING USE CAUSING ADDITIONAL GALLING BETWEEN THE BUR AND COLLET THAN DURING TYPICAL USE AND IS LIKELY TO RESULT IN A SEIZURE OF THE BUR IN THE COLLET AND OTHER FAILURES. IT IS POSSIBLE THAT THE USER EXPERIENCED DIFFICULTY ACHIEVING FULL INSERTION OF THE BUR AND PROPER LOCK DOWN INSIDE THE COLLET DUE TO METAL TO METAL GALLING BETWEEN THE BUR/CUTTER, THE DRILL COLLET NOSE HOUSING, AND/OR THE COLLET SPINDLE DURING A PRIOR OPERATION. EXCESSIVE BENDING, LATERAL FORCES, EXCEEDING THE RECOMMENDED PSI SETTING, AND OTHER AGGRESSIVE USE MAY CONTRIBUTE TO THE GALLING CONDITION.

Description of Event or Problem · 1

BURR MOVED IN THE HANDPIECE DURING MILLING OF THE BONE CAUSING A LARGER AMOUNT OF BONE TO BE REMOVED IN THE FEMUR. THE BURR THEN BECAME STUCK AND UNABLE TO REPOSITION OR REMOVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
655587 ZIMMER MILL SURGICAL MILLING HANDPIECE HSZ BRASSELER USA, MEDICAL L.L.C. DEMO ZIMMER MILL 00-5927-040-00D 08010094

Patients

Seq Age Sex Outcome Treatment
1