FDA Adverse Event Malfunction Summary report: N

PARADIGM INSULIN INFUSION PUMP

MDR report key: 5130613 · Received October 6, 2015

Report

Report Number
2032227-2015-53603
Event Type
Malfunction
Date Received
October 6, 2015
Date of Event
September 18, 2015
Report Date
September 18, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
K040676
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Additional Manufacturer Narrative · 1

ALL BUTTONS FUNCTIONED PROPERLY. HOWEVER, FOUND CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS WERE NOTED DURING TESTING. THE PUMP WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON THE DISPLAY WINDOW AND A CRACKED CASE AT THE DISPLAY WINDOW CORNER.

Description of Event or Problem · 1

IT WAS REPORTED THAT CUSTOMER RECEIVED A BUTTON ERROR ALARM. INSULIN PUMP WOULD NEED TO BE REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
661574 PARADIGM INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-715WWS

Patients

Seq Age Sex Outcome Treatment
1