IMMULITE 2000 XPI
Report
- Report Number
- 2247117-2015-00053
- Event Type
- Malfunction
- Date Received
- October 5, 2015
- Date of Event
- June 15, 2015
- Report Date
- October 13, 2015
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- JJQ
- PMA / PMN Number
- P010053/S007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
INITIAL MDR 2247117-2015-00053 WAS FILED ON 10/52015. ADDITIONAL INFORMATION (10/13/2015): THE CUSTOMER CONTACTED SIEMENS CUSTOMER CARE CENTER (CCC). THE CUSTOMER INDICATED THAT BASED ON INTERFERENCE TESTING FOR HETEROPHILIC ANTIBODIES THE RESULTS HAVE BEEN DISCUSSED WITH THE LAB MANAGER WHO HAS ACCEPTED THE EXPLANATION FOR THE DISCORDANT LOW CALCITONIN RESULT. THE CUSTOMER HAS ALSO INDICATED THAT THE CALCITONIN ASSAY IS PERFORMING AS PER SPECIFICATIONS AND DOES NOT NEED ANY FURTHER ASSISTANCE ON THIS ISSUE. THE SYSTEM IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
THE CAUSE OF THE FALSELY LOW CALCITONIN RESULT ON THE IMMULITE 2000 XPI INSTRUMENT IS UNKNOWN. SIEMENS HEALTHCARE IS INVESTIGATING THE ISSUE.
THE CUSTOMER HAS OBTAINED A FALSELY LOW RESULT ON PATIENT SAMPLE FOR THE CALCITONIN ASSAY ON AN IMMULITE 2000 XPI. THE SAME SAMPLE WAS RUN ON AN ALTERNATE PLATFORM AND THE RESULT WAS HIGHER. THE INITIAL RESULT WAS REPORTED AND QUESTIONED BY THE PHYSICIAN (S) AS IT DID NOT FIT THE CLINICAL PICTURE. THE RESULT ON THE ALTERNATE PLATFORM WAS REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY LOW RESULT OBTAINED ON THE IMMULITE 2000 XPI.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 657630 | IMMULITE 2000 XPI | IMMULITE 2000 XPI | JJQ | SIEMENS HEALTHCARE DIAGNOSTICS INC. | IMMULITE 2000 XPI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |