FDA Adverse Event Malfunction Summary report: N

ALERE CHOLESTECH LDX MULTIANALYTE CONTROLS

MDR report key: 5125624 · Received October 2, 2015

Report

Report Number
2027969-2015-00825
Event Type
Malfunction
Date Received
October 2, 2015
Date of Event
September 4, 2015
Report Date
September 4, 2015
Manufacturer
ALERE SAN DIEGO, INC.
Product Code
JJY
PMA / PMN Number
K102700
Removal / Correction Number
2027969-08/25/15-001C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
DIETICIAN

Narratives

Additional Manufacturer Narrative · 1

MDR IS BEING FILED DUE TO FIELD ACTION: REF-2027969-08/25/15-001C. INVESTIGATION PENDING.

Additional Manufacturer Narrative · 1

CONTROL LOT C3091A WAS ASSOCIATED WITH FIELD ACTION PER INTERNAL (B)(4) FOR INCORRECT EXPECTED VALUE CARD PROVIDED WITH THIS CONTROL LOT WITH THE INCORRECT CONCENTRATION ASSIGNMENT RANGES FOR THE TOTAL CHOLESTEROL AND TRIGLYCERIDES ANALYTES. INVESTIGATION WILL BE PERFORMED UNDER INTERNAL (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED THAT CONTROLS WERE OUTSIDE OF THE EXPECTED RANGE - HIGH. RESULTS WERE AS FOLLOWS: TC HIGH TWO TIMES FOR L1. TC=160, 157 (108-152); TRG=132 (129-191). THE RESULTS WERE WITHIN THE CORRECTED EXPECTED VALUES. TECHNICAL SERVICE REVIEWED FACTORS AND INFORMED CUSTOMER OF NOTIFICATION; PROVIDED CORRECTED EXPECTED VALUES RANGES AND EMAILED THE NOTIFICATION TO THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
656651 ALERE CHOLESTECH LDX MULTIANALYTE CONTROLS MULTI-ANALYTE CONTROLS JJY ALERE SAN DIEGO, INC. 88773 C3091A

Patients

Seq Age Sex Outcome Treatment
1