ALERE CHOLESTECH LDX MULTIANALYTE CONTROLS
Report
- Report Number
- 2027969-2015-00825
- Event Type
- Malfunction
- Date Received
- October 2, 2015
- Date of Event
- September 4, 2015
- Report Date
- September 4, 2015
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- JJY
- PMA / PMN Number
- K102700
- Removal / Correction Number
- 2027969-08/25/15-001C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- DIETICIAN
Narratives
MDR IS BEING FILED DUE TO FIELD ACTION: REF-2027969-08/25/15-001C. INVESTIGATION PENDING.
CONTROL LOT C3091A WAS ASSOCIATED WITH FIELD ACTION PER INTERNAL (B)(4) FOR INCORRECT EXPECTED VALUE CARD PROVIDED WITH THIS CONTROL LOT WITH THE INCORRECT CONCENTRATION ASSIGNMENT RANGES FOR THE TOTAL CHOLESTEROL AND TRIGLYCERIDES ANALYTES. INVESTIGATION WILL BE PERFORMED UNDER INTERNAL (B)(4).
CUSTOMER REPORTED THAT CONTROLS WERE OUTSIDE OF THE EXPECTED RANGE - HIGH. RESULTS WERE AS FOLLOWS: TC HIGH TWO TIMES FOR L1. TC=160, 157 (108-152); TRG=132 (129-191). THE RESULTS WERE WITHIN THE CORRECTED EXPECTED VALUES. TECHNICAL SERVICE REVIEWED FACTORS AND INFORMED CUSTOMER OF NOTIFICATION; PROVIDED CORRECTED EXPECTED VALUES RANGES AND EMAILED THE NOTIFICATION TO THE CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 656651 | ALERE CHOLESTECH LDX MULTIANALYTE CONTROLS | MULTI-ANALYTE CONTROLS | JJY | ALERE SAN DIEGO, INC. | 88773 | C3091A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |