FDA Adverse Event
Malfunction
Summary report: N
STEALTHSTATION TREON
MDR report key: 512461
·
Received February 6, 2004
Report
- Report Number
- MW1031130
- Event Type
- Malfunction
- Date Received
- February 6, 2004
- Date of Event
- February 4, 2004
- Report Date
- February 6, 2004
- Manufacturer
- MEDTRONIC SURGICAL NAVIGATION TECHNOLOGIES
- Product Code
- HAW
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Description of Event or Problem · 1
POWER CORD ON THE STEALTHSTATION TREON WAS PARTIALLY CUT THROUGH BY THE METALLIC STRAIN RELIEF ON THE CORD RETRACTING MECHANISM. THIS RESULTED IN A SHORT BETWEEN THE POWER LINE HOT AND THE CHASSIS OF THE MACHINE. POTENTIAL FOR ELECTRIC SHOCK, FIRE, OR LOSS OF POWER TO CRITICAL EQUIPMENT. NO INJURY REPORTED. BEGINNING OF SURGERY WAS DELAYED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STEALTHSTATION TREON | SURGICAL NAVIGATION SYSTEM | HAW | MEDTRONIC SURGICAL NAVIGATION TECHNOLOGIES | STEALTHSTATION TREON | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|