FDA Adverse Event Malfunction Summary report: N

STEALTHSTATION TREON

MDR report key: 512461 · Received February 6, 2004

Report

Report Number
MW1031130
Event Type
Malfunction
Date Received
February 6, 2004
Date of Event
February 4, 2004
Report Date
February 6, 2004
Manufacturer
MEDTRONIC SURGICAL NAVIGATION TECHNOLOGIES
Product Code
HAW
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

POWER CORD ON THE STEALTHSTATION TREON WAS PARTIALLY CUT THROUGH BY THE METALLIC STRAIN RELIEF ON THE CORD RETRACTING MECHANISM. THIS RESULTED IN A SHORT BETWEEN THE POWER LINE HOT AND THE CHASSIS OF THE MACHINE. POTENTIAL FOR ELECTRIC SHOCK, FIRE, OR LOSS OF POWER TO CRITICAL EQUIPMENT. NO INJURY REPORTED. BEGINNING OF SURGERY WAS DELAYED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STEALTHSTATION TREON SURGICAL NAVIGATION SYSTEM HAW MEDTRONIC SURGICAL NAVIGATION TECHNOLOGIES STEALTHSTATION TREON *

Patients

Seq Age Sex Outcome Treatment
1 *