FDA Adverse Event Malfunction Summary report: N

SIGMA HP UNI ANTERIOR CHISEL

MDR report key: 5124404 · Received October 2, 2015

Report

Report Number
1818910-2015-32009
Event Type
Malfunction
Date Received
October 2, 2015
Date of Event
September 24, 2015
Report Date
September 24, 2015
Manufacturer
DEPUY ORTHOPAEDICS, INC.
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CONCLUSION AND JUSTIFICATION STATUS FOR MDR: EXAMINATION OF THE SUBMITTED SIGMA HP UNI ANTERIOR CHISEL CONFIRMED THE CHISEL BLADE IS BROKEN INTO TWO SECTIONS. THERE IS DAMAGE TO THE BLADE TIP INDICATING REPEATED CONTACT WITH THE CUTTING BLOCK FIXATION PINS. THE TIP DAMAGE IS DUE A PIN BEING PRESENT IN THE TRACK CLOSEST TO THE PATELLA WHILE CHISELING. THE SURGICAL TECHNIQUE STATES " THE CHISEL IS REQUIRED ONLY IN THE TRACK CLOSEST TO THE PATELLA. PIN THE HOLE FARTHEST AWAY FROM THE PATELLA AND USE THE ANTERIOR CHISEL IN THE TRACK CLOSEST TO THE PIN. THE ROOT CAUSE IS USER ERROR. BASED ON THE SURGICAL TECHNIQUE NOTING THE RECOMMENDED PIN LOCATION WHILE USING THE CHISEL, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY. (B)(4).

Description of Event or Problem · 1

WHILE TAPPING IN CHISEL INTO FEMORAL CUT GUIDE THE CHISEL BROKE IN TWO PIECES. NO SURGICAL DELAY OCCURRED AND NO PIECE OF INSTRUMENTATION WAS LEFT IN THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
654818 SIGMA HP UNI ANTERIOR CHISEL KNEE INSTRUMENT/TRIAL LXH DEPUY ORTHOPAEDICS, INC. PG1009

Patients

Seq Age Sex Outcome Treatment
1