FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 5124008 · Received October 2, 2015

Report

Report Number
2032227-2015-52133
Event Type
Malfunction
Date Received
October 2, 2015
Date of Event
September 14, 2015
Report Date
June 28, 2016
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED STUCK IN THE MOTOR ERROR LOOP DURING BOLUS AND BASAL DELIVERY. UNABLE TO CONFIRM E70 ERROR ALARM DUE TO ERASED HISTORY FILE. THE MOTOR WAS TESTED OUTSIDE THE DEVICE AND PASSED; THE MOTOR MAY HAVE HAD AN INTERMITTENT FAILURE THAT WAS NOT DETECTED DURING OUR TESTING. UNABLE TO PERFORM A21 ERROR TEST DUE TO MOTOR ERROR LOOP. HOWEVER, THE INSULIN PUMP WAS RECEIVED WITH OPERATING CURRENTS WITHIN SPECIFICATIONS AND UNIT PASSED SELF-TEST, REWIND, BASIC OCCLUSION TEST, PRIME TEST AND DISPLACEMENT TEST. THE INSULIN PUMP HAD BROKEN BATTERY TUBE THREADS, CRACKED CASE AT THE DISPLAY WINDOW CORNERS AND MINOR SCRATCHES ON THE LCD WINDOW.

Description of Event or Problem · 1

IT WAS REPORTED OF A MOTOR ERROR FROM THE INSULIN PUMP. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS UNKNOWN. THE CUSTOMER WAS ADVISED THAT IT COULD BE CONNECTION ISSUES. THE PUMP WILL BE REPLACED.