FDA Adverse Event Injury Summary report: N

CODMAN VPV SYSTEM

MDR report key: 5117993 · Received October 1, 2015

Report

Report Number
1226348-2015-10572
Event Type
Injury
Date Received
October 1, 2015
Manufacturer
CODMAN & SHURTLEFF, INC. / MEDOS S.A.
Product Code
JXG
PMA / PMN Number
PK061876
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

510K: K061876 & K050739. (B)(4). UPON COMPLETION OF THE INVESTIGATION A FOLLOW UP REPORT WILL BE FILED.

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS EVALUATED BY THE PRODUCT MANAGER AND WAS FOUND TO BE FUNCTIONING PROPERLY. A REVIEW OF MANUFACTURING RECORDS WAS PERFORMED AND NO DISCREPANCIES WERE NOTED WHEN THE INSTRUMENT WAS RELEASED TO STOCK. THE ISSUE REPORTED BY THE CUSTOMER COULD NOT BE REPRODUCED. TRENDS WILL BE MONITORED FOR THIS OR SIMILAR COMPLAINTS. AT PRESENT, WE CONSIDER THIS COMPLAINT TO BE CLOSED.

Description of Event or Problem · 1

OVER THE PREVIOUS 2 WEEKS, THE PROGRAMMER IS CONSISTENTLY FAILING TO RECOGNIZE THE REPROGRAMMING, NECESSITATING AN XRAY FOR VERIFICATION. AT PRESENT, XRAYS INDICATE THE PROGRAMMER IS EFFECTIVELY CHANGING TO THE DESIRED PRESSURE SETTING, HOWEVER, THE ACOUSTIC VERIFICATION CONSISTENTLY FAILS TO INDICATE THAT THE SHUNT SETTING HAS BEEN CHANGED. NO OUTPATIENT CONSEQUENCES, BUT PATIENTS WOULD HAVE TO BE SENT TO XRAY FOR VERIFICATION PLACING ADDITIONAL COSTS ON HOSPITAL. UPON REPS INVESTIGATION, IT WAS NOTED THAT THE ACOUSTIC PRONG, LOCATED IN THE CENTRE OF PROGRAMMER HANDPIECE, IS STUCK IN A DEPRESSED POSITION, PREVENTING IT FROM COMING IN CONTACT WITH PATIENT SKIN. OVER 5 INSTANCES BETWEEN AUGUST 17, 2015 AND SEPTEMBER 4, 2015. NO SPECIFIC DATES PROVIDED. REP WILL BRING REPLACEMENT DEVICE TO ACCOUNT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
652059 CODMAN VPV SYSTEM VALVE PROGRAMMER JXG CODMAN & SHURTLEFF, INC. / MEDOS S.A. CKLCD1

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention