FDA Adverse Event Malfunction Summary report: N

RESTORE

MDR report key: 5117688 · Received October 1, 2015

Report

Report Number
3004209178-2015-19160
Event Type
Malfunction
Date Received
October 1, 2015
Date of Event
September 9, 2015
Report Date
September 9, 2015
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LGW
PMA / PMN Number
P840001
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

CONCOMITANT PRODUCTS: PRODUCT ID 3708340, SERIAL # (B)(4), IMPLANTED: (B)(6) 2006, PRODUCT TYPE EXTENSION, PRODUCT ID 3708340, SERIAL # (B)(4), IMPLANTED: (B)(6) 2006, PRODUCT TYPE EXTENSION; PRODUCT ID 3487A-33, LOT # V000574, IMPLANTED: (B)(6) 2006, PRODUCT TYPE LEAD; PRODUCT ID 3487A-33, LOT # V001185, IMPLANTED: (B)(6) 2006, PRODUCT TYPE LEAD.

Description of Event or Problem · 1

THE HEALTH CARE PROFESSIONAL (HCP) REPORTED THAT THERE WERE OUT OF RANGE (OOR) IMPEDANCES WHEN TESTING AT 0.75V. THE FOLLOWING: 02, 04, 05, 06, 07, 12, 13, 14, 15 WERE OOR. IT WAS NOTED THAT THE REASON FOR THE INQUIRY WAS TO GET AN ESTIMATE ON BATTERY LIFE AS THEY WOULD BE REPLACING THE IMPLANTABLE NEUROSTIMULATOR (INS) DUE TO NORMAL BATTERY CIRCUMSTANCES. SETTINGS PROVIDED WERE: 1, 5.0V, 360USEC, 75HZ. ON (B)(6) 2015, THE HCP REPORTED THAT THE OOR IMPEDANCES WERE DUE TO SOMEONE ENTERING THE NUMBERS IN THE CLINICIAN PROGRAMMER WRONG. ONCE IT WAS CORRECTED THERE WAS STILL 1 CONTACT THAT WAS OUT OF RANGE. IT WAS BELIEVED THAT CONTACT #1, BUT THE CAUSE FOR THIS WAS NOT DETERMINED. IT WAS THOUGHT EVERYTHING WAS RESOLVED AT THE TIME AND THERE WERE NO PATIENT SYMPTOMS OR FURTHER ISSUES NOTED. OTHER RELEVANT MEDICAL HISTORY INCLUDES FAILED BACK SURGERY SYNDROME. NO INTERVENTIONS OR OUTCOME WERE REPORTED REGARDING THIS EVENT. FURTHER FOLLOW-UP IS BEING CONDUCTED TO OBTAIN THIS INFORMATION. IF ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP REPORT W ILL BE SENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
650334 RESTORE STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF LGW MEDTRONIC MED REL MEDTRONIC PUERTO RICO 37711

Patients

Seq Age Sex Outcome Treatment
1 00064 YR