FDA Adverse Event Malfunction Summary report: N

CERAMIC ELECTRODE TIP J-HK F/GK372R

MDR report key: 5114972 · Received September 30, 2015

Report

Report Number
2916714-2015-00845
Event Type
Malfunction
Date Received
September 30, 2015
Report Date
May 20, 2016
Manufacturer
AESCULAP AG
Product Code
GEI
PMA / PMN Number
K970541
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

MANUFACTURING SITE EVALUATION: EVALUATION ON-GOING.

Additional Manufacturer Narrative · 1

ACCORDING TO THE AVAILABLE INFORMATION, THERE WERE NO NEGATIVE CONSEQUENCES FOR PATIENT. THE EQUIPMENT ARRIVED WITHOUT VISUAL DEFECTS. TEST AND ANALYSIS PERFORMED: MICROSCOPE AND DIGITAL CAMERA A VISIBLE INSPECTION OF THE HANDLE. IT LOOKS NEW WITHOUT ANY TRACES OF USAGE. THE CLAMPING MECHANISM FOR THE ELECTRODES WORKED WELL. THE CONNECTION PIN IN THE HANDLE PIECE IS FIX AND NOT BENT. THE ELECTRICAL RESISTANCE PIN-WORKING END IS LOWER THAN 1 AND THEREFORE IT IS OKAY. NEITHER HANDLE PIECE NOR THE SHAFT EXHIBIT ANY HINTS FOR INSULATION FAULTS OR DISRUPTIVE BREAKDOWNS. A VISIBLE INSPECTION OF THE CORD (GK246). AS WELL AS THE HANDLE, THE CABLE SHOWS NO VISIBLE TRACES OF USAGE. NO CRACKS OR PRESSURE MARKS. NEITHER THE PLUG NOR THE SOCKET EXHIBIT DAMAGED AREAS. THE ELECTRICAL RESISTANCE OF THE CORD IS LOWER THAN 1 AND IS THERE OKAY. NEITHER DURING BENDING THE CONNECTION AREA OF THE PLUG NOR THE BENDING OF THE AREA OF THE SOCKET CREATING ANY CONTACT INTERRUPTIONS. THE LAST TEST WAS THE RF-HOOK ELECTRODE. IT EXHIBITS NO SIGNS OF WEAR AND TEAR, NO BENDING OR INSULATION DEFECTS. THE PRODUCT DOES NOT REQUIRE BATCH MANAGEMENT; A REVIEW OF THE DEVICE QUALITY AND MANUFACTURING HISTORY RECORDS IS NOT POSSIBLE. NONE OF THE COMPLAINED PRODUCTS SHOWS ANY DEFECT OR DEVIATION. THERE IS NO DEFECT IN THE PRODUCTS AND NO DESCRIPTION IN DETAIL WHAT KIND OF PROBLEM OCCURRED, IT IS NOT POSSIBLE TO DETERMINE A ROOT CAUSE. ACCORDING TO SA-(B)(4) THERE IS NO CAPA NECESSARY.

Description of Event or Problem · 1

COUNTRY OF COMPLAINT: (B)(6). INCIDENT INVOLVING RENU SET 4, A BOWEL INJURY OCCUR DURING A LAPAROSCOPIC SURGERY. COMPONENTS REPORTED- HANDLE F/MONOPOLAR ELECTRODES 5MM_PRODUCT # GK372R. MONOPOLAR CABLE W/LG.PIN 12FEET_PRODUCT # GK246.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
645720 CERAMIC ELECTRODE TIP J-HK F/GK372R OTHER INSTRUMENTS GEI AESCULAP AG GK383R

Patients

Seq Age Sex Outcome Treatment
1 Other