CERAMIC ELECTRODE TIP J-HK F/GK372R
Report
- Report Number
- 2916714-2015-00845
- Event Type
- Malfunction
- Date Received
- September 30, 2015
- Report Date
- May 20, 2016
- Manufacturer
- AESCULAP AG
- Product Code
- GEI
- PMA / PMN Number
- K970541
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
MANUFACTURING SITE EVALUATION: EVALUATION ON-GOING.
ACCORDING TO THE AVAILABLE INFORMATION, THERE WERE NO NEGATIVE CONSEQUENCES FOR PATIENT. THE EQUIPMENT ARRIVED WITHOUT VISUAL DEFECTS. TEST AND ANALYSIS PERFORMED: MICROSCOPE AND DIGITAL CAMERA A VISIBLE INSPECTION OF THE HANDLE. IT LOOKS NEW WITHOUT ANY TRACES OF USAGE. THE CLAMPING MECHANISM FOR THE ELECTRODES WORKED WELL. THE CONNECTION PIN IN THE HANDLE PIECE IS FIX AND NOT BENT. THE ELECTRICAL RESISTANCE PIN-WORKING END IS LOWER THAN 1 AND THEREFORE IT IS OKAY. NEITHER HANDLE PIECE NOR THE SHAFT EXHIBIT ANY HINTS FOR INSULATION FAULTS OR DISRUPTIVE BREAKDOWNS. A VISIBLE INSPECTION OF THE CORD (GK246). AS WELL AS THE HANDLE, THE CABLE SHOWS NO VISIBLE TRACES OF USAGE. NO CRACKS OR PRESSURE MARKS. NEITHER THE PLUG NOR THE SOCKET EXHIBIT DAMAGED AREAS. THE ELECTRICAL RESISTANCE OF THE CORD IS LOWER THAN 1 AND IS THERE OKAY. NEITHER DURING BENDING THE CONNECTION AREA OF THE PLUG NOR THE BENDING OF THE AREA OF THE SOCKET CREATING ANY CONTACT INTERRUPTIONS. THE LAST TEST WAS THE RF-HOOK ELECTRODE. IT EXHIBITS NO SIGNS OF WEAR AND TEAR, NO BENDING OR INSULATION DEFECTS. THE PRODUCT DOES NOT REQUIRE BATCH MANAGEMENT; A REVIEW OF THE DEVICE QUALITY AND MANUFACTURING HISTORY RECORDS IS NOT POSSIBLE. NONE OF THE COMPLAINED PRODUCTS SHOWS ANY DEFECT OR DEVIATION. THERE IS NO DEFECT IN THE PRODUCTS AND NO DESCRIPTION IN DETAIL WHAT KIND OF PROBLEM OCCURRED, IT IS NOT POSSIBLE TO DETERMINE A ROOT CAUSE. ACCORDING TO SA-(B)(4) THERE IS NO CAPA NECESSARY.
COUNTRY OF COMPLAINT: (B)(6). INCIDENT INVOLVING RENU SET 4, A BOWEL INJURY OCCUR DURING A LAPAROSCOPIC SURGERY. COMPONENTS REPORTED- HANDLE F/MONOPOLAR ELECTRODES 5MM_PRODUCT # GK372R. MONOPOLAR CABLE W/LG.PIN 12FEET_PRODUCT # GK246.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 645720 | CERAMIC ELECTRODE TIP J-HK F/GK372R | OTHER INSTRUMENTS | GEI | AESCULAP AG | GK383R |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |