FDA Adverse Event Malfunction Summary report: N

SOVEREIGN COMPACT CONSOLE

MDR report key: 5112773 · Received September 30, 2015

Report

Report Number
3006695864-2015-00661
Event Type
Malfunction
Date Received
September 30, 2015
Date of Event
September 3, 2015
Report Date
February 11, 2016
Manufacturer
ABBOTT MEDICAL OPTICS
Product Code
HQC
PMA / PMN Number
K060366
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 1

(B)(4). (B)(6). DEVICE MANUFACTURE DATE: THE INFORMATION WAS REQUESTED AND HAS NOT YET BEEN RECEIVED. THE FIELD SERVICE SPECIALIST (FSS) VISITED CUSTOMER SITE TO INSPECT THE UNIT. THE FSS CHECKED THE SYSTEM AND FOUND BATTERY LEAKAGE IN SINGLE BOARD COMPUTER (SBC) PRINTED CIRCUIT BOARD (PCB); THEREFORE, THE SINGLE BOARD COMPUTER PCB WAS REPLACED. FSS CHECKED ALL PARAMETERS AND FUNCTION AS PER ABBOTT MEDICAL OPTICS (AMO) CHECKLIST. THE UNIT WAS FOUND WORKING FINE AND IT MET AMO SPECIFICATIONS. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT MEDICAL OPTICS HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 1

DEVICE MANUFACTURE DATE: 10/2/2009. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT MEDICAL OPTICS HAS BEEN SUBMITTED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED NO ALTERNATING CURRENT (AC), THAT THE SYSTEM WOULD NOT TURN ON AND THAT SMOKE CAME OUT OF THE MACHINE. THERE WAS NO PATIENT INVOLVEMENT REPORTED AND NO PATIENT TREATMENTS DELAYED OR CANCELLED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
645818 SOVEREIGN COMPACT CONSOLE PHACOFRAGMENTATION UNIT HQC ABBOTT MEDICAL OPTICS CMP680300

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown