SOVEREIGN COMPACT CONSOLE
Report
- Report Number
- 3006695864-2015-00661
- Event Type
- Malfunction
- Date Received
- September 30, 2015
- Date of Event
- September 3, 2015
- Report Date
- February 11, 2016
- Manufacturer
- ABBOTT MEDICAL OPTICS
- Product Code
- HQC
- PMA / PMN Number
- K060366
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
(B)(4). (B)(6). DEVICE MANUFACTURE DATE: THE INFORMATION WAS REQUESTED AND HAS NOT YET BEEN RECEIVED. THE FIELD SERVICE SPECIALIST (FSS) VISITED CUSTOMER SITE TO INSPECT THE UNIT. THE FSS CHECKED THE SYSTEM AND FOUND BATTERY LEAKAGE IN SINGLE BOARD COMPUTER (SBC) PRINTED CIRCUIT BOARD (PCB); THEREFORE, THE SINGLE BOARD COMPUTER PCB WAS REPLACED. FSS CHECKED ALL PARAMETERS AND FUNCTION AS PER ABBOTT MEDICAL OPTICS (AMO) CHECKLIST. THE UNIT WAS FOUND WORKING FINE AND IT MET AMO SPECIFICATIONS. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT MEDICAL OPTICS HAS BEEN SUBMITTED.
DEVICE MANUFACTURE DATE: 10/2/2009. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT MEDICAL OPTICS HAS BEEN SUBMITTED.
THE CUSTOMER REPORTED NO ALTERNATING CURRENT (AC), THAT THE SYSTEM WOULD NOT TURN ON AND THAT SMOKE CAME OUT OF THE MACHINE. THERE WAS NO PATIENT INVOLVEMENT REPORTED AND NO PATIENT TREATMENTS DELAYED OR CANCELLED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 645818 | SOVEREIGN COMPACT CONSOLE | PHACOFRAGMENTATION UNIT | HQC | ABBOTT MEDICAL OPTICS | CMP680300 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |