FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 5112106 · Received September 29, 2015

Report

Report Number
1720753-2015-03961
Event Type
Malfunction
Date Received
September 29, 2015
Date of Event
August 20, 2015
Report Date
September 29, 2015
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
OWB
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE GENERATOR DRIVE BOARD WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED AN AD CHANNEL 15 ERROR AND WOULD NOT BOOT UP. THERE WAS NO PATIENT INJURY OR DEATH REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
641620 9800 FLUOROSCOPIC X-RAY OWB GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1