FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 5112106
·
Received September 29, 2015
Report
- Report Number
- 1720753-2015-03961
- Event Type
- Malfunction
- Date Received
- September 29, 2015
- Date of Event
- August 20, 2015
- Report Date
- September 29, 2015
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- OWB
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE GENERATOR DRIVE BOARD WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED AN AD CHANNEL 15 ERROR AND WOULD NOT BOOT UP. THERE WAS NO PATIENT INJURY OR DEATH REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 641620 | 9800 | FLUOROSCOPIC X-RAY | OWB | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |