FDA Adverse Event Other Summary report: N

SC7000

MDR report key: 511044 · Received January 14, 2004

Report

Report Number
1220063-2004-00001
Event Type
Other
Date Received
January 14, 2004
Date of Event
December 12, 2003
Report Date
January 13, 2004
Manufacturer
DRAEGER MEDICAL SYSTEMS, INC.
Product Code
DSI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

IT WAS REPORTED: IN THE EMERGENCY ROOM, A NURSE WAS ATTEMPTING TO DOCK THE PT MONITOR ONTO THE DOCKING STATION AND IT WAS NOTED THAT THE MONITOR FELL OFF OF THE DOCKING STATION AND HIT THE FLOOR. THERE WAS NO INJURY INVOLVED. A BIOMED AT THE FACILITY PERFORMED AN EVALUATION OF THE DOCKING STATION AND REPORTED THAT THE CONNECTION DOOR COVER / LOCKING MECHANISM ON THE DOCKING STATION WAS LOOSE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SC7000 MODULAR PT MONITOR DSI DRAEGER MEDICAL SYSTEMS, INC. 5202994E539U NA

Patients

Seq Age Sex Outcome Treatment
1 NA Other DRAEGER MEDICAL SYSTEMS - INFINITY DOCKING STATION| (PN 5594887EE56U).