FDA Adverse Event Malfunction Summary report: N

COLLEAGE 3

MDR report key: 510835 · Received February 3, 2004

Report

Report Number
MW1031016
Event Type
Malfunction
Date Received
February 3, 2004
Date of Event
December 4, 2003
Report Date
December 31, 2003
Manufacturer
BAXTER PRODUCT SERVICE MGR
Product Code
FRN
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
IA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

IV PUMP FOUND TO BE UNDER DELIVERING GREATER THAN 5% OF THE TIME. REPORTER HAS FOUND BY RANDOM SAMPLE OF CHECKING THESE PUMPS, FIVE PUMPS NOW ARE UNDER DELIVERING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGE 3 IV PUMP FRN BAXTER PRODUCT SERVICE MGR * *
2 COLLEAGE 3 IV PUMP FRN BAXTER PRODUCT SERVICE MGR * *

Patients

Seq Age Sex Outcome Treatment
1 NA Other