FDA Adverse Event
Summary report: N
RAPIDCLEAN(R)
MDR report key: 5107878
·
Received September 18, 2015
Report
- Report Number
- 3007208013-2015-00022
- Date Received
- September 18, 2015
- Date of Event
- July 14, 2015
- Report Date
- September 18, 2015
- Manufacturer
- SYMMETRY SURGICAL INC.
- Product Code
- HTX
- UDI-DI
- 00887482102527
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE RAPIDCLEAN CERAMIC COATED CONVERTIBLE MICRO KERRISON RONGEUR SCREW/PIN THAT SITS IN THE TOP PORTION OF THE KERRISON CAME APART FROM THE MAIN SHAFT. CONTRACT MANUFACTURER EVALUATED THE DEVICE INDICATING: THE SCREW CAME LOOSE, CAUSING MALFUNCTION. DEVICE WAS REPAIRED; SCREW WAS SECURED. CONTRACT MANUFACTURER INDICATED; THROUGH INADEQUATE MAINTENANCE OF THE INSTRUMENT MAY CAUSE THE SCREW LOCK TO SWITCH OFF. PROPER INSTRUCTIONS FOR USE OF MAINTAINING INSTRUMENTS (IFU) ARE BEING PROVIDED TO THE CUSTOMER. REPORT RECEIVED 17-SEPT-2015.
Description of Event or Problem · 1
THE SCREW/PIN THAT SITS IN THE TOP PORTION OF THE KERRISON CAME APART FROM THE MAIN SHAFT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 619405 | RAPIDCLEAN(R) | CERAMIC COATED CONVERTIBLE MICRO KERRISON RONGEUR | HTX | SYMMETRY SURGICAL INC. | 58-3200J-BD | BJ1113 | 00887482102527 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |