FDA Adverse Event Summary report: N

RAPIDCLEAN(R)

MDR report key: 5107878 · Received September 18, 2015

Report

Report Number
3007208013-2015-00022
Date Received
September 18, 2015
Date of Event
July 14, 2015
Report Date
September 18, 2015
Manufacturer
SYMMETRY SURGICAL INC.
Product Code
HTX
UDI-DI
00887482102527
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE RAPIDCLEAN CERAMIC COATED CONVERTIBLE MICRO KERRISON RONGEUR SCREW/PIN THAT SITS IN THE TOP PORTION OF THE KERRISON CAME APART FROM THE MAIN SHAFT. CONTRACT MANUFACTURER EVALUATED THE DEVICE INDICATING: THE SCREW CAME LOOSE, CAUSING MALFUNCTION. DEVICE WAS REPAIRED; SCREW WAS SECURED. CONTRACT MANUFACTURER INDICATED; THROUGH INADEQUATE MAINTENANCE OF THE INSTRUMENT MAY CAUSE THE SCREW LOCK TO SWITCH OFF. PROPER INSTRUCTIONS FOR USE OF MAINTAINING INSTRUMENTS (IFU) ARE BEING PROVIDED TO THE CUSTOMER. REPORT RECEIVED 17-SEPT-2015.

Description of Event or Problem · 1

THE SCREW/PIN THAT SITS IN THE TOP PORTION OF THE KERRISON CAME APART FROM THE MAIN SHAFT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
619405 RAPIDCLEAN(R) CERAMIC COATED CONVERTIBLE MICRO KERRISON RONGEUR HTX SYMMETRY SURGICAL INC. 58-3200J-BD BJ1113 00887482102527

Patients

Seq Age Sex Outcome Treatment
1