FDA Adverse Event Injury Summary report: N

40 OS CHAIR

MDR report key: 5106118 · Received September 25, 2015

Report

Report Number
1523574-2015-00083
Event Type
Injury
Date Received
September 25, 2015
Date of Event
September 14, 2015
Report Date
November 5, 2015
Manufacturer
FERNO-WASHINGTON, INC.
Product Code
FPP
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
EMERGENCY MEDICAL TECHNICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. THE CONDITION OF THE UNIT INDICATES IT HAS HAD EXTENSIVE USE FOR ITS AGE. THE LOCKING MECHANISM WAS FUNCTIONAL AND NO MALFUNCTIONS WERE DETECTED THROUGHOUT THE EVALUATION. IT COULD NOT BE DETERMINED HOW THE LOCK BAR WAS BENT; HOWEVER, IT WAS CONFIRMED THE BEND TO THE LOCK BAR DOES NOT IMPAIR THE FUNCTION OF THE LOCK BAR. THE CUSTOMER WAS ASKED IF THEY HAD RECEIVED ANY FEEDBACK FROM THE HOSPITAL OR THE PATIENT REGARDING SERIOUS INJURY RELATED TO THE INCIDENT AND THEY STATED THEY HAD RECEIVED NO SUCH COMMUNICATION.

Description of Event or Problem · 1

IT WAS REPORTED WHILE MOVING A PATIENT ACROSS A DOORWAY SADDLE, THE STAIR CHAIR ALLEGEDLY GAVE WAY AND THE EMT LOST THEIR FOOTING ALLOWING THE CHAIR AND PATIENT TO FALL BACKWARDS. THE PATIENT ALLEGEDLY HIT HER HEAD ON THE O2 BAG AND LATER COMPLAINED OF SOME HEAD PAIN. THE CUSTOMER REPORTED THE ONLY NOTICABLE DAMAGE TO THE CHAIR WAS A BENT RELEASE BAR; HOWEVER, THE CHAIR IS BEING RETURNED TO FERNO FOR EVALUATION.

Description of Event or Problem · 1

IT WAS REPORTED WHILE MOVING A PATIENT ACROSS A DOORWAY SADDLE, THE STAIR CHAIR ALLEGEDLY GAVE WAY AND THE EMT LOST THEIR FOOTING ALLOWING THE CHAIR AND PATIENT TO FALL BACKWARDS. THE PATIENT ALLEGEDLY HIT HER HEAD ON THE O2 BAG AND LATER COMPLAINED OF SOME HEAD PAIN. THE CUSTOMER REPORTED THE ONLY NOTICABLE DAMAGE TO THE CHAIR WAS A BENT RELEASE BAR; HOWEVER, THE CHAIR IS BEING RETURNED TO FERNO FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
635266 40 OS CHAIR 40 OS CHAIR FPP FERNO-WASHINGTON, INC. 0714851

Patients

Seq Age Sex Outcome Treatment
1 57 YR Other