530G INSULIN PUMP
Report
- Report Number
- 2032227-2015-48414
- Event Type
- Malfunction
- Date Received
- September 24, 2015
- Date of Event
- September 4, 2015
- Report Date
- September 4, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- PMA / PMN Number
- 120010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).
THE UP ARROW BUTTON DID NOT RESPOND DUE TO A FLATTENED BUTTON DOME SWITCH. NO UNLOCKED KEYPAD CONNECTOR ON THE LCD BOARD NOTED. THE PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW, AND A CRACKED RESERVOIR TUBE LIP.
CUSTOMER REPORTED VIA PHONE CALL TO HAVE EXPERIENCED KEYPAD ANOMALY. CUSTOMER WAS NOT ABLE TO CHANGE INFUSION SET DUE TO THE ESC BUTTON NOT RESPONDING. CUSTOMER'S BLOOD GLUCOSE WAS 280 MG/DL. CUSTOMER DOES NOT KNOW WHAT CAUSED ANOMALY. AFTER TROUBLESHOOTING, CUSTOMER WAS ADVISED THAT INSULIN PUMP WOULD NEED TO BE REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 631700 | 530G INSULIN PUMP | OZO | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-751NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 26 YR |