FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 5102136 · Received September 24, 2015

Report

Report Number
2032227-2015-48414
Event Type
Malfunction
Date Received
September 24, 2015
Date of Event
September 4, 2015
Report Date
September 4, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Additional Manufacturer Narrative · 1

THE UP ARROW BUTTON DID NOT RESPOND DUE TO A FLATTENED BUTTON DOME SWITCH. NO UNLOCKED KEYPAD CONNECTOR ON THE LCD BOARD NOTED. THE PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW, AND A CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

CUSTOMER REPORTED VIA PHONE CALL TO HAVE EXPERIENCED KEYPAD ANOMALY. CUSTOMER WAS NOT ABLE TO CHANGE INFUSION SET DUE TO THE ESC BUTTON NOT RESPONDING. CUSTOMER'S BLOOD GLUCOSE WAS 280 MG/DL. CUSTOMER DOES NOT KNOW WHAT CAUSED ANOMALY. AFTER TROUBLESHOOTING, CUSTOMER WAS ADVISED THAT INSULIN PUMP WOULD NEED TO BE REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
631700 530G INSULIN PUMP OZO OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-751NAS

Patients

Seq Age Sex Outcome Treatment
1 26 YR