FDA Adverse Event
Malfunction
Summary report: N
COBAS MIRA
MDR report key: 510207
·
Received January 12, 2004
Report
- Report Number
- 1823260-2004-00008
- Event Type
- Malfunction
- Date Received
- January 12, 2004
- Date of Event
- August 18, 2003
- Report Date
- August 19, 2003
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJF
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- UNKNOWN
Narratives
Description of Event or Problem · 1
ORIGINAL GLUCOSE RESULT FOR A PT SAMPLE WAS 90 MG/DL. WHEN RETESTED, THE RESULT WAS 331 MG/DL. FIELD SERVICE REP REPLACED THE SAMPLE NEEDLE AND SYRINGE TIP TO REPAIR THE ANALYZER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COBAS MIRA | CLINICAL CHEMISTRY ANALYZER | JJF | ROCHE DIAGNOSTICS | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Other |