FDA Adverse Event Injury Summary report: N

WATERPIK WATER FLOSSER

MDR report key: 5101391 · Received September 24, 2015

Report

Report Number
0001712259-2015-00006
Event Type
Injury
Date Received
September 24, 2015
Date of Event
September 23, 2015
Report Date
November 19, 2015
Manufacturer
WATER PIK, INC.
Product Code
EFS
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UNIT IN QUESTION HAS NOT YET BEEN RETURNED FOR TESTING AND EVALUATION.

Additional Manufacturer Narrative · 1

RETURNED UNIT TESTED AND FOUND TO BE WITHIN THE NORMAL OPERATING PRESSURE RANGE AND WAS OPERATING AS DESIGNED. THERE IS NO KNOWN SITUATION WHERE PRESSURE CAN TEMPORARILY INCREASE. MOST LIKELY SITUATION IS THAT CUSTOMER AIMED WATER JET AT SOFT TISSUE INSTEAD OF GUM AREA AND CAUSED PAIN. THIS IS WARNED AGAINST IN THE PRODUCT INSTRUCTIONS FOR USE.

Description of Event or Problem · 1

PER THE CUSTOMER....FATHER CLAIMS THE (B)(6) DAUGHTER INJURED THE ROOF OF HER MOUTH WHILE FLOSSING. SAYS DAUGHTER HAS USED HER WP-360 FOR SOME TIME WITH THE HIGH PRESSURE TIP AND HAS HAD NO PROBLEMS. SUDDENLY THE PRESSURE BECAME MUCH HIGHER THAN USUAL AND CAUSED SMALL ABRASIONS ON THE ROOF OF HER MOUTH. DAUGHTER HAS BRACES, DID NOT RECEIVE MEDICAL ATTENTION AND ABRASIONS HEALED QUICKLY. CUSTOMER REQUESTED A REPLACEMENT UNIT.

Description of Event or Problem · 1

PER THE CUSTOMER....FATHER CLAIMS THE (B)(6) YEAR OLD DAUGHTER INJURED THE ROOF OF HER MOUTH WHILE FLOSSING. SAYS DAUGHTER HAS USED HER WP-360 FOR SOME TIME WITH THE HIGH PRESSURE TIP AND HAS HAD NO PROBLEMS. SUDDENLY THE PRESSURE BECAME MUCH HIGHER THAN USUAL AND CAUSED SMALL ABRASIONS ON THE ROOF OF HER MOUTH. DAUGHTER HAS BRACES, DID NOT RECEIVE MEDICAL ATTENTION AND ABRASIONS HEALED QUICKLY. CUSTOMER REQUESTED A REPLACEMENT UNIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
631981 WATERPIK WATER FLOSSER ORAL IRRIGATOR EFS WATER PIK, INC. WP-360 KL120316GA

Patients

Seq Age Sex Outcome Treatment
1 11 YR Other