BLOOD GLUCOSE SYSTEM
Report
- Report Number
- 1052693-2015-01783
- Event Type
- Malfunction
- Date Received
- September 24, 2015
- Date of Event
- August 27, 2015
- Report Date
- December 31, 2015
- Manufacturer
- NIPRO DIAGNOSTICS, INC.
- Product Code
- NBW
- PMA / PMN Number
- K080641
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA
- Reporter Occupation
- OTHER
Narratives
(B)(4). PRODUCT EVALUATION IN PROCESS.
(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER HAD AN INACCURATE REFERENCE.
CUSTOMER COMPLAINS OF LOW RESULTS. CUSTOMER STATES THAT HE FEELS PHYSICALLY FINE AND DENIES THE NEED FOR MEDICAL ATTENTION. THE CUSTOMER'S EXPECTED BLOOD RESULTS ARE 90-120MG/DL FASTING. VERIFIED TEST STRIPS EXPIRE 06/16/2018. CONFIRMED CUSTOMER'S TEST STRIPS ARE BEING STORED PROPERLY, AND OPENED VIAL DATE IS 08/24/2015. CUSTOMER PERFORMED A BACK TO BACK BLOOD TEST 158MG/DL AND 158MG/DL BOTH TIMES. REVIEWED METER MEMORY (B)(6). CUSTOMER IS COMPARING HIS TRUERESULT METER TO HIS BROTHERS TRUERESULT METER. HE RAN A BLOOD GLUCOSE TEST ON (B)(6) 2015 @9AM (FASTING) ON HIS TRUERESULT METER HE GOT 55MG/DL AND HIS BROTHERS TRUERESULT METER GAVE HIM 79MG/DL. NO ADVERSE EVENT REPORTED.
CUSTOMER COMPLAINS OF LOW RESULTS. CUSTOMER STATES THAT HE FEELS PHYSICALLY FINE AND DENIES THE NEED FOR MEDICAL ATTENTION. THE CUSTOMER'S EXPECTED BLOOD RESULTS ARE 90-120MG/DL FASTING. VERIFIED TEST STRIPS EXPIRE 06/16/2018. CONFIRMED CUSTOMER'S TEST STRIPS ARE BEING STORED PROPERLY, AND OPENED VIAL DATE IS (B)(6) 2015. CUSTOMER PERFORMED A BACK TO BACK BLOOD TEST 158MG/DL AND 158MG/DL BOTH TIMES. REVIEWED METER MEMORY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 631773 | BLOOD GLUCOSE SYSTEM | NBW | NIPRO DIAGNOSTICS, INC. | TRUERESULT | REPLACED F |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |