FDA Adverse Event Malfunction Summary report: N

QUANTUM Q6 EDGE WITH TB3 POWER SEATING SYSTEM

MDR report key: 5100572 · Received September 24, 2015

Report

Report Number
5100572
Event Type
Malfunction
Date Received
September 24, 2015
Date of Event
July 15, 2015
Report Date
August 25, 2015
Manufacturer
PRIDE MOBILITY PRODUCTS CORP
Product Code
IOR
Adverse Event
Yes
Report Source
User Facility report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

PATIENT'S MOTHER REPORTED HEARING "SOME NOISE" WHILE UTILIZING THE WHEELCHAIR THAT DAY. AT THE END OF THE DAY AFTER THE PATIENT WAS TRANSFERRED OUT OF THE CHAIR AND STORED IN THE ROOM FOR CHARGING, THEY HEARD A LOUD NOISE AND UPON CHECKING THE ROOM WITH THE CHAIR THE BACK REST UPPER SYSTEM HAD FALLEN OFF. NO INJURY(S). THE PATIENT HAS A BACKUP CHAIR. THE MANUFACTURER IS REPLACING THE ENTIRE CHAIR AND SEATING AT NO COST. MANUFACTURER RESPONSE FOR WHEELCHAIR, QUANTUM Q6 EDGE WITH TB3 POWER SEATING SYSTEM (PER SITE REPORTER): WORKING WITH THE MANUFACTURER ON RESOLUTION AND ACTION PLAN FOR THIS AND ALL OTHER CHAIRS WE HAVE DISTRIBUTED. THE PERMANENT FIX WILL BE REPLACING ALL RAILS PROVIDED BETWEEN A FIFTEEN MONTH PERIOD WITH THE NEW DESIGNED RAIL SYSTEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
632348 QUANTUM Q6 EDGE WITH TB3 POWER SEATING SYSTEM WHEELCHAIR, POWERED IOR PRIDE MOBILITY PRODUCTS CORP Q6 EDGE

Patients

Seq Age Sex Outcome Treatment
1 16 YR NO