BLOOD GLUCOSE SYSTEM
Report
- Report Number
- 1052693-2015-01777
- Event Type
- Malfunction
- Date Received
- September 23, 2015
- Date of Event
- September 1, 2015
- Report Date
- December 11, 2015
- Manufacturer
- NIPRO DIAGNOSTICS, INC.
- Product Code
- NBW
- PMA / PMN Number
- K080641
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA
- Reporter Occupation
- OTHER
Narratives
(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND. THE MOST LIKELY UNDERLYING ROOT CAUSE OF THIS MALFUNCTION WAS IDENTIFIED AS A STRIP ISSUE.
(B)(4). PRODUCT EVALUATION IN PROCESS.
CUSTOMER COMPLAINS OF LO RESULTS. CUSTOMER STATES HE FEELS WELL AND STATES HE DOES NOT REQUIRE MEDICAL ATTENTION. THE CUSTOMER'S EXPECTED BLOOD RESULTS ARE 120-130MG/DL. VERIFIED THE STRIPS EXPIRE 01/16/2018. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WERE FIRST OPENED IN (B)(6) 2015. REVIEWED METER MEMORY: LO 09/01/2015 07:00:00 AM FASTING:YES 127MG/DL, (B)(6) 2015 07:00:00 AM, FASTING: NO. 125MG/DL, (B)(6) 2015 07:00:00 AM, FASTING: YES. 101MG/DL, (B)(6) 2015 07:00:00 AM, FASTING: YES. 113MG/DL, (B)(6) 2015 07:00:00 AM, FASTING: YES. CUSTOMERS CONCERN: LO (B)(6) 2015 7:00AM. NO ADVERSE EVENT REPORTED.
CUSTOMER COMPLAINS OF LO RESULTS. CUSTOMER STATES HE FEELS WELL AND STATES HE DOES NOT REQUIRE MEDICAL ATTENTION. THE CUSTOMER'S EXPECTED BLOOD RESULTS ARE 120-130MG/DL. VERIFIED THE STRIPS EXPIRE 01/16/2018. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WERE FIRST OPENED IN (B)(6) 2015. REVIEWED METER MEMORY: (B)(6). CUSTOMERS CONCERN: LO (B)(6) 2015 7:00AM. NO ADVERSE EVENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 629686 | BLOOD GLUCOSE SYSTEM | NBW | NIPRO DIAGNOSTICS, INC. | TRUERESULT | PS2201 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |