FDA Adverse Event Malfunction Summary report: N

BLOOD GLUCOSE SYSTEM

MDR report key: 5098955 · Received September 23, 2015

Report

Report Number
1052693-2015-01777
Event Type
Malfunction
Date Received
September 23, 2015
Date of Event
September 1, 2015
Report Date
December 11, 2015
Manufacturer
NIPRO DIAGNOSTICS, INC.
Product Code
NBW
PMA / PMN Number
K080641
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND. THE MOST LIKELY UNDERLYING ROOT CAUSE OF THIS MALFUNCTION WAS IDENTIFIED AS A STRIP ISSUE.

Additional Manufacturer Narrative · 1

(B)(4). PRODUCT EVALUATION IN PROCESS.

Description of Event or Problem · 1

CUSTOMER COMPLAINS OF LO RESULTS. CUSTOMER STATES HE FEELS WELL AND STATES HE DOES NOT REQUIRE MEDICAL ATTENTION. THE CUSTOMER'S EXPECTED BLOOD RESULTS ARE 120-130MG/DL. VERIFIED THE STRIPS EXPIRE 01/16/2018. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WERE FIRST OPENED IN (B)(6) 2015. REVIEWED METER MEMORY: LO 09/01/2015 07:00:00 AM FASTING:YES 127MG/DL, (B)(6) 2015 07:00:00 AM, FASTING: NO. 125MG/DL, (B)(6) 2015 07:00:00 AM, FASTING: YES. 101MG/DL, (B)(6) 2015 07:00:00 AM, FASTING: YES. 113MG/DL, (B)(6) 2015 07:00:00 AM, FASTING: YES. CUSTOMERS CONCERN: LO (B)(6) 2015 7:00AM. NO ADVERSE EVENT REPORTED.

Description of Event or Problem · 1

CUSTOMER COMPLAINS OF LO RESULTS. CUSTOMER STATES HE FEELS WELL AND STATES HE DOES NOT REQUIRE MEDICAL ATTENTION. THE CUSTOMER'S EXPECTED BLOOD RESULTS ARE 120-130MG/DL. VERIFIED THE STRIPS EXPIRE 01/16/2018. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WERE FIRST OPENED IN (B)(6) 2015. REVIEWED METER MEMORY: (B)(6). CUSTOMERS CONCERN: LO (B)(6) 2015 7:00AM. NO ADVERSE EVENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
629686 BLOOD GLUCOSE SYSTEM NBW NIPRO DIAGNOSTICS, INC. TRUERESULT PS2201

Patients

Seq Age Sex Outcome Treatment
1 0 YR