BLOOD GLUCOSE SYSTEM
Report
- Report Number
- 1052693-2015-01773
- Event Type
- Malfunction
- Date Received
- September 23, 2015
- Date of Event
- August 31, 2015
- Report Date
- December 11, 2015
- Manufacturer
- NIPRO DIAGNOSTICS, INC.
- Product Code
- NBW
- PMA / PMN Number
- K080641
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER
Narratives
(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: THE USER HAD AN INACCURATE REFERENCE.
(MANUFACTURER NARRATIVE = T, CORRECTED DATA = F) INTERNAL REPORT # (B)(4). PRODUCT NOT YET RETURNED FOR EVALUATION.
CUSTOMER COMPLAINS OF LOW RESULTS. CUSTOMER STATES THEY FEEL WELL AND THAT NO MEDICAL ATTENTION. THE CUSTOMER'S EXPECTED BLOOD RESULTS ARE 166MG/DL. VERIFIED THE STRIPS EXPIRE 07/21/2018. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WERE FIRST OPENED (B)(6) 2015. CUSTOMER PERFORMED BACK TO BACK BLOOD TEST 84MG/DL AND 77MG/DL FASTING. REVIEWED METER MEMORY: 1:95MG/DL (B)(6) 2015 04:18:00 PM FASTING:YES. 2:100MG/DL (B)(6) 2015 04:14:00 PM FASTING:YES. 3:71MG/DL (B)(6) 2015 08:50:00 AM FASTING:YES. 4:72MG/DL (B)(6) 2015 05:44:00 PM FASTING:YES. 5:120MG/DL (B)(6) 2015 04:08:00 PM FASTING:YES. CUSTOMERS CONCERN:71 MG/DL AND 72 MG/DL. NO ADVERSE EVENT REPORTED.
CUSTOMER COMPLAINS OF LOW RESULTS. CUSTOMER STATES THEY FEEL WELL AND THAT NO MEDICAL ATTENTION. THE CUSTOMER'S EXPECTED BLOOD RESULTS ARE 166MG/DL. VERIFIED THE STRIPS EXPIRE 07/21/2018. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WERE FIRST OPENED (B)(6) 2015. CUSTOMER PERFORMED BACK TO BACK BLOOD TEST 84MG/DL AND 77MG/DL FASTING. REVIEWED METER MEMORY: 95MG/DL, (B)(6) 2015, 04:18:00 PM, FASTING:YES. 100MG/DL, (B)(6) 2015, 04:14:00 PM, FASTING:YES. 71MG/DL, (B)(6) 2015, 08:50:00 AM, FASTING:YES. 72MG/DL, (B)(6) 2015, 05:44:00 PM, FASTING:YES. 120MG/DL, (B)(6) 2015, 04:08:00 PM ,FASTING:YES. CUSTOMERS CONCERN:71 MG/DL AND 72 MG/DL. NO ADVERSE EVENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 628748 | BLOOD GLUCOSE SYSTEM | NBW | NIPRO DIAGNOSTICS, INC. | TRUERESULT | PS2416 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |