FDA Adverse Event Malfunction Summary report: N

ACCESS CEA

MDR report key: 5093797 · Received September 22, 2015

Report

Report Number
2122870-2015-00591
Event Type
Malfunction
Date Received
September 22, 2015
Date of Event
August 25, 2015
Report Date
August 25, 2015
Manufacturer
BECKMAN COULTER
Product Code
DHX
PMA / PMN Number
K031270
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS SUCH AS AGE, DATE OF BIRTH, SEX, AND WEIGHT. (B)(6). A BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER'S SITE TO ASSESS THE INSTRUMENT'S PERFORMANCE. THE FSE PERFORMED PRESSURE SENSOR CALIBRATION; NO HARDWARE MALFUNCTIONS WERE IDENTIFIED THAT MAY HAVE CAUSED, OR CONTRIBUTED, TO THIS EVENT. THE IMPLICATED REAGENT PACK WAS DISCARDED BY THE CUSTOMER AND IS NOT AVAILABLE FOR INVESTIGATION. INVESTIGATION OF THE EVENT DETERMINED THAT AN ERROR OCCURRED ON THE MANUFACTURER'S REAGENT PACK FILL LINE. DUE TO THE ERROR, PARTIALLY FILLED REAGENT PACKS WERE RELEASED FOR USE. WHEN USED TO ANALYZE PATIENT SAMPLES, PARTIALLY FILLED REAGENT PACKS COULD CAUSE NO VALUE/QSD-FLAGGED RESULTS. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED OBTAINING TWO (2) NO VALUE/QSD-FLAGGED (INSUFFICIENT SAMPLE OR REAGENT VOLUME DISPENSED) CARCINOEMBRYONIC ANTIGEN (ACCESS CEA) PATIENT RESULTS ON THE LABORATORY'S UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM, SERIAL NUMBER (B)(4). THE CUSTOMER DID NOT INDICATE IF THESE SAMPLES WERE REANALYZED AND WHAT RESULTS THEY OBTAINED IF REANALYSIS DID OCCUR. UPON VISUAL INSPECTION, THERE WERE NO SIGNS OF REAGENT PACK LEAKAGE. NONE OF THE QSD-FLAGGED ACCESS CEA RESULTS WERE REPORTED OUTSIDE THE LABORATORY. THERE WAS NO CHANGE IN PATIENT TREATMENT ASSOCIATED WITH THIS EVENT. CALIBRATION AND QUALITY CONTROL (QC) DATA WERE NOT PROVIDED FOR THIS EVENT. THE CUSTOMER DID NOT PROVIDE SPECIFIC DETAILS REGARDING THE PATIENT SAMPLES OR SAMPLE PROCESSING. THERE WAS NO INDICATION OF SAMPLE INTEGRITY ISSUES RELATED TO THIS EVENT. A BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ASSESS THE INSTRUMENT'S PERFORMANCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
626576 ACCESS CEA SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN, PRODUCT CODE: DHX DHX BECKMAN COULTER NA 595027

Patients

Seq Age Sex Outcome Treatment
1