FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 5092969 · Received September 21, 2015

Report

Report Number
2531779-2015-33795
Event Type
Malfunction
Date Received
September 21, 2015
Report Date
August 31, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 10/21/2105 DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2015 WITH THE FOLLOWING FINDINGS: LOSS OF PRIME WARNINGS WERE OBSERVED IN THE PUMP'S BLACK BOX WHERE THE FORCE READINGS DID NOT REACH ZERO. THE REWIND, LOAD CARTRIDGE, AND PRIME STEPS WERE PERFORMED SUCCESSFULLY WITH NO ALARMS. A FORCE SENSOR CALIBRATION TEST CONFIRMED THAT THE SENSOR WAS DETECTING THE CORRECT FORCE. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO LOSS OF PRIMES OCCURRING. THE PUMP WAS OPENED AND NO DAMAGE WAS FOUND TO THE FORCE SENSOR CIRCUIT.

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS BUT NOT YET EVALUATED. WHEN THE DEVICE EVALUATION IS COMPLETE, A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THERE WERE MULTIPLE LOSS OF PRIME WARNINGS WITH DIFFERENT CARTRIDGES. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
623815 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 9 YR