CHROMID CPS ELITE TRANSLUCENT 100 PLT
Report
- Report Number
- 3002769706-2015-00096
- Event Type
- Malfunction
- Date Received
- September 21, 2015
- Report Date
- August 27, 2015
- Manufacturer
- BIOMERIEUX SA
- Product Code
- JXA
- PMA / PMN Number
- CL I, EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
DEVICE NOT RETURNED TO MANUFACTURER.
REVIEW OF QC BATCH RECORDS INDICATE CONFORMANCE TO SPECIFICATIONS. DURING THE INVESTIGATION, SIXTY (60) CHROMID TM CPS ELITE AGAR PLATES WERE TESTED; THE PLATES WERE ASSOCIATED WITH THREE (3) DIFFERENT TIMEFRAMES OF THE MANUFACTURING PROCESS FOR THE IDENTIFIED BATCH (11:25, 16:24 AND 19:40). THE PLATES WERE NOT INOCULATED. THE PLATES WERE INCUBATED FOR FIVE (5) DAYS AT 33-37ºC. RESULTS CONFORMED TO SPECIFICATIONS FOR CONTAMINATION; ZERO (0) CONTAMINATIONS OBSERVED IN 60 PLATES. THERE IS NO EVIDENCE OF A CONTAMINATION ISSUE WITH THE CHROMID TM CPS ELITE AGAR PRODUCT.
A CUSTOMER CALLED IN COMPLAINING OF A CONTAMINATE ON (B)(4),CHROMID(TM) CHROMID CPS PLATES, LOT 1004145580 DISCOVERED AFTER INCUBATION. BACTERIUM CONTAMINATION IS PRESENTING AS A BROWNISH-YELLOW UNIDENTIFIED COLONY. CUSTOMER REPORTS NO PATIENT RESULTS WERE AFFECTED. AN INTERNAL INVESTIGATION HAS BEEN LAUNCHED USING RETAINED PLATES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 623545 | CHROMID CPS ELITE TRANSLUCENT 100 PLT | CHROMID CPS ELITE TRANSLUCENT 100 PLT | JXA | BIOMERIEUX SA | 1004145580 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |