FDA Adverse Event
Injury
Summary report: N
BIRD 6400ST
MDR report key: 50914
·
Received July 9, 1996
Report
- Report Number
- 50914
- Event Type
- Injury
- Date Received
- July 9, 1996
- Date of Event
- May 20, 1996
- Report Date
- May 23, 1996
- Manufacturer
- BIRD PRODUCTS CORP.
- Product Code
- CBK
- Report Source
- User Facility report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
VENTILATOR ALARMED AND THEN DISPLAYED INOPERATIVE READING. PT BAGGED AND 100% O2 SATURATION MAINTAINED. VENTILATOR EVALUATED AND FOUND TO HAVE DIAPHRAM AND EXHALATION VALVE STICKING. PROBLEM CORRECTED, VENTILATOR REINSTATED. NO HARM TO PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BIRD 6400ST | VENTILATOR | CBK | BIRD PRODUCTS CORP. | 6400ST(9611) | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 28 YR | Life Threatening| R |